Run
module
type
Run module
name
Configuration
23 hours throughput
[1]
Performance
Cap.
length
(cm)
Pol.
type
Run
time
(min)
3500
(8-cap.)
3500xL
(24-
cap.)
Range
[2]
Sizing Precision
[3]
50bp-
400bp
401bp-
600bp
601bp-
1200bp
HID
HID36_POP4
HID36_POP4xl
36
POP-4
™
£
35
³
312
³
936
£
60
to
³
400
<0.15
NA
[4]
NA
[4]
SNaPshot
™
SNaPshot50_PO
P7
SNaPshot50_PO
P7xl
50
POP-7
™
£
30
³
376
³
1104
£
40
to
³
120
<0.50
NA
[4]
NA
[4]
[1]
Throughput (Samples / Day): The total number of samples run in 23 hours (0.5 hour for user interaction and 0.5 hour for warm-up time).
[2]
Resolution Range: The range of bases over which the resolution (peak spacing interval divided by the peak width at half-max in a GS600 or
GS1200 LIZ size standard sample sized with a third order fit) is
³
1. The table shows the resolution range in
³
90% of samples.
[3]
Sizing Precision: Standard deviation of sizes for one allele in the DS-33 install standard sized with the GS600 LIZ size standard across multiple
capillaries in the same run. For one injection to pass, 100% of the alleles in that injection must meet the intra-run sizing precision specifications.
The table shows the sizing precision of 100% of alleles in
³
90% of samples.
[4]
Not applicable because of the size of the fragments collected in the run.
Dye sets
Table 7
Sequence analysis dye sets
Dye Set
Application Name
E (v1.1 BigDye
™
Terminator)
Rapid DNA sequencing
Z (v3.1 BigDye
™
Terminator)
DNA sequencing
Table 8
Fragment analysis dye sets
Application
E5
SNaPshot
™
kit
G5
DNA sizing for 5-dye chemistry
J6
DNA sizing for 6-dye chemistry
F
DNA sizing for 4-dye chemistry
D
DNA sizing for 4-dye chemistry
AnyDye
DNA sizing
Sequencing
analysis dye sets
Fragment analysis
dye sets for all
applications
Appendix B
Run modules and dye sets
Dye sets
B
284
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1