Thermal Cycler Temperature Verification System For GeneAmp
®
PCR Systems 2720, 9600, 9700, 9800,
and Veriti
®
System User Guide
Chapter 7
Temperature non-uniformity test
3. Press
F4
(Util).
The Utilities screen opens.
4. Press
F1
(Diag). The Diagnostics screen opens.
5. Press
F3
(TmpVer). The Temperature Verification screen opens.
6. Press
F2
(TNU) to automatically configure the GeneAmp
®
9700 for the
temperature non-uniformity test. The TNU Performance screen opens.
below.
Running the test
The temperature non-uniformity test uses the 0.5-mL probe assembly to test the
temperature uniformity of 12 different wells in the sample block.
Note:
If necessary, press
F5
(Cancel) to exit the test.
To run the temperature non-uniformity test:
“Setting up the 0.5-mL probe assembly”
above.
2. Press
F1
(Run) to start the temperature non-uniformity test. The TNU
Performance screen opens with the setpoint value displayed.
Note:
The sample block must be within
±
1.0 °C of the setpoint. In addition, the
cover must be within
±
1 °C of 105 °C. It may take several minutes for the
GeneAmp
®
9700 to ramp up.
Temperature Verification
Temp - Calibration Verification
TNU - Temperature Non-Uniformity
F1
F3
F4
F5
F2
Temp
Exit
TNU
TNU Performance
Sample temp = xx.x°C Cover temp = xxx°C
Place Probe in well A1
Press Run
F1
F3
F4
F5
F2
Run
Cancel
TNU Performance
Sample temp = xx.x°C Cover temp = xxx°C
Setpoint is 94°C
Sample must be within 1.0°C of setpoint
F1
F3
F4
F5
F2
Cancel