AT851 

Battery Aging Tester

]

 

User’s Guide

길  우  트  레  이  딩

서울특별시 영등포구 선유로33길 15
703호(양평동3가, 에이스테크노타워)<150-962>

02)2636-0009  팩스02)2636-4753

www.gilwoo.co.kr   [email protected]
www.gilwooshop.co.kr

Содержание AT851

Страница 1: ...RE REVISIONS This manual applies directly to instruments that have the firmware Rev A AT851 Battery Aging Tester User s Guide 길 우 트 레 이 딩 서울특별시 영등포구 선유로33길 15 703호 양평동3가 에이스테크노타워 150 962 02 2636 0009 팩스02 2636 4753 www gilwoo co kr giltron naver com www gilwooshop co kr ...

Страница 2: ...mer s failure to comply with these requirements Ground The Instrument To avoid electric shock hazard the instrument chassis and cabinet must be connected to a safety earth ground by the supplied power cable with earth blade DO NOT Operate In An Explosive Atmosphere Do not operate the instrument in the presence of inflammable gasses or fumes Operation of any electrical instrument in such an environ...

Страница 3: ... product must be returned to a service facility designated by Applent The buyer shall prepay shipping charges to Applent and Applent shall pay shipping charges to return the product to the Buyer However the Buyer shall pay all shipping charges duties and taxes for products returned to Applent from another country Applent warrants that its software and firmware designated by Applent for use with an...

Страница 4: ...er Current OC 16 3 9 3 Over Power OP 16 3 9 4 Reverse Voltage RV 16 3 9 5 Over Heat OH 17 4 Basic Operation 18 4 1 Modes of Operation 18 4 1 1 Constant Current Mode CC 18 4 1 2 Constant Voltage Mode CV 19 4 1 3 Constant Power Mode CP 19 4 1 4 Constant Resistance Mode CR 20 4 2 Functions of Operation 21 4 3 Setup the Maximum Current I max 21 4 4 Setup the Maximum Voltage V max 22 4 5 Setup the Maxi...

Страница 5: ... 20 Figure 4 6 Functions of Operation 21 Figure 4 7 I max Setup Page 21 Figure 4 8 Von off Operation 22 Figure 4 9 V ON Setup Page 23 Figure 4 10 V OFF Setup Page 23 Figure 4 11 Load ON Timer Setup Page 23 Figure 4 12 File Operation Page 24 Figure 5 8 Short Circuit Testing Page 26 Figure 5 9 Set up Discharge Current 27 Figure 5 10 Set up cut off voltage 27 Figure 5 11 Battery Testing Function Page...

Страница 6: ...port do not turn on the power switch in case that get an electrical shock Check whether the packing box or shock absorbing material used packaging the instrument has been damaged Referring to Packing List in the packing box check whether all packaged items supplied together with the meter have been provided as per the specified optioned 1 2 Setting up Fuse Figure 1 1 Fuse Holder Line 110VAC 220V A...

Страница 7: ...on condensation Altitude 0 to 2 000m Vibration Max 0 5 G 5 Hz to 500 Hz 1 4 Cleaning To prevent electrical shock disconnect the AT851 power cable from the receptacle before cleaning Use dry cloths or cloths slightly dipped in water to clean the case Do not attempt to clean the internal of the AT851 WARNING Don t Use Organic Solvents such as alcohol or gasoline to clean the Instrument 1 5 How to Re...

Страница 8: ...8 AT851 Battery Aging Tester Remove Handle Lift the handle perpendicular to the unit while pulling it in the direction of 1 ...

Страница 9: ...ANCE POWER 0 150W AT8511 0 300W AT8512 Accuracy 1 CONSTANCE RESISTANCE 0 4000Ω Accuracy 1 Battery Aging Test Discharge Charge Rest Battery Aging Charge and Rest Timer 9999 minutes Battery Capacity 999 99AH Battery Discharge Timer 999H Maximum Loop Times 999 times 2 2 Features VFD Display Operation Mode NORMAL CC CV CP CR TRANSIENT TRN SEQUENCE LIST SEQ AUTOMATIC TESTING FUNCTION ATF BATTERY TESTIN...

Страница 10: ...1 Front panel Figure 3 1 Front panel Table 3 1 Front panel description No Name and Description 1 POWER SWITCH Push down ON Push up OFF 2 VFD Displaying measurement results instrument status and user s interface menus 3 Knob To choose menu item and input number 4 Terminals 5 Keypad II Multi function keys Numeric and 2nd Function 6 Shift Pressing this key before using 2nd Function 7 Keypad I Dual Fu...

Страница 11: ...ion Mode SEQ Automatic Testing Function ATF Battery Testing Function BAT Short Circuit Operation SHT 4 5 EX TRIG TRIG BUS Trigger Indicator EX TRIG External Trigger Indicator See also Trigger 6 FMS Rate Indicator F Fast M Medium S Slow 7 MAX Maximum Value 8 MIN Minimum Value 9 Shift Lit when Shift key pressed 10 File No 11 DELAY Load on Timer Function start working 12 VLMT Display when Von Voff Op...

Страница 12: ... CR will be showed on Display 4 Shift V On Off Set up Von value and Voff value for Von Off operation VLMT indicator will be showed on VFD File File Save Read Erase Operation The FILE setup page will be displayed List Sequence List Operation The LIST setup page will be displayed Rotary knob to adjust the value in Test and Setup mode Rotary knob to choose menu in menu mode Choose menu item Esc Escap...

Страница 13: ...rations WARNING FIRE HAZARD To satisfy safety requirements load wires must be heavy enough to not overheated while carrying the short circuit output current of the device connects to AT851 Input connections are made to the and binding posts on the front panel A major consideration in making input connections is the wire size The minimum wire size required to prevent overheating may not be large en...

Страница 14: ...f AT851 are connected to the output of the power supply Remote sensing compensates for the voltage drop in applications require long lead lengths Before using remote sensing connections you must set the MENU SENSE to ON state A 4W indicator will be displayed on VFD Figure 3 7 The remote sensing interface on real panel GND TRIG SENSE TRIGGER Figure 3 8 Remote Sensing ...

Страница 15: ...C SOWER SOURCE OBAI PANBI 3 7 Turning the Input On off Figure 3 9 Input On Off switchers on front panel The Input can be toggled on and off by pressing ESC ON OFF key The Input state ON or OFF will be displayed on display 3 Figure 3 9 The Input State 3 8 Operation Range ...

Страница 16: ... display 3 In the CV mode transition mode and list mode the buzzer beeps and OC will been displayed once input current exceeds the 102 I MAX 3 9 3 Over Power OP CV mode If the input power exceeds the power limit 101 P MAX load will work in the over power protection state The buzzer beeps current value will be flashed and OP will be displayed If the input power exceeds the power limit 102 P MAX loa...

Страница 17: ...17 Start up 3 9 5 Over Heat OH Once the load internal power devices temperature exceeds the safe limits 80 C load input will turn off OH will be displayed on VFD and buzzer beeps ...

Страница 18: ...ad will sink a current in accordance with the programmed value regardless of the input voltage See following figure Figure 4 1 Constant Current Mode Setting Current Loed Currirt I Loed Irtut Voltegi V AT851 provides two methods to set the current value by using the numeric keypad and the knob Method 1 Setup the current value by using the numeric keypad Table 4 1 Setup input current I Set by using ...

Страница 19: ...and the knob Method 1 Setup the voltage value by using the numeric keypad Table 4 2 Setup voltage V Set by using numeric keypad Procedure Operation details Step 1 Press V Set If the mode is not the CV mode press V Set again Step 2 Enter a new value using numeric keypad or rotary the knob to adjust the value Example press key 1 0 5 Step 3 Press Enter to confirm this input Method 2 Setup the voltage...

Страница 20: ...3 Press Enter to confirm this input Method 2 Setup the power value by using knob The value can be continually changed from the previews value according the knob rotation The knob rotation magnification can be set in Menu Knob item 4 1 4 Constant Resistance Mode CR In CR mode AT851 will sink a current linearly proportional to the input voltage in accordance with the programmed resistance According ...

Страница 21: ... of Operation AT851 includes six functions of operation 1 Normal Operation CC CV CP CR 2 Transient Operation TRN 3 Sequence List Operation SEQ 4 Automatic Testing Function ATF 5 Battery Testing Function BAT 6 Short Circuit Testing Function SHT Figure 4 5 Functions of Operation Normal Operation CC CV CP CR Transient Operation TRN Sequence List Operation SEQ Automatic Testing Function ATF Battery Te...

Страница 22: ... are used to input the P max value 1st The value can be continually changed from the previews value according the knob rotation The knob rotation magnification can be set in Menu Knob item 2nd Using numeric keypad to input value 4 6 Von off Operation You can set voltage value Von off to control the input state on off for electronic load When the input voltage reaches the Von value the load s input...

Страница 23: ...n off the input automatically Load ON Timer Setup Procedure Press Shift Timer to enter Load ON Timer setup Page Figure 4 10 Load ON Timer Setup Page Enter the timer Value by using numeric keypad or knob Enter 0 can disable Load ON Timer function The unit is second and maximum time is 60000s 4 8 File Operation Saving Loading Erasing Settings You can save these following values into internal non vol...

Страница 24: ...r the load input turned ON the display 3 s measurement result can be switched by or key Current Meas Value I Power Meas Value P Resistance Meas Value R Battery Meas Capacity B Availability in Battery Testing Function Battery discharge time T Availability in Battery Testing Function 4 10 Menu The menu settings will be saved into internal non volatile memory automatically These Settings will be load...

Страница 25: ...ate F indicator will be showed on VFD TRIG Trigger Source INT Internal Trigger BUS BUS Trigger EXT External Trigger Trigger signal inputs from handler interface on the real panel COMM Communication 232 RS 232C 485 RS 485 Optional BAUD Baud Rate Select 4800 9600 19200 38400 57600 115 2k ADDR RS 485 Address Optional 0 254 INIT Return to factory default value OFF ON ...

Страница 26: ...n mode CV CC CP CR and the load input is turned off choose an operation mode CV CC CP or CR Step 2 press shift Short to enter Short Circuit Testing Function Figure 5 1 Short Circuit Testing Page Step 3 Turn on load input by ON OFF Step 4 You can press or key to watch Current I Power P or Resistance R measurement value Step 5 Turn off load input by pressing ON OFF key Step 6 Press shift Short to re...

Страница 27: ... Page During the testing press to watch current I power P battery capacity B or discharge time T The discharge time format is HHH MM hour minute When the battery voltage drops below the setting value load input turns off Step 7 Press shift Battery to return normal operation mode 5 3 Battery Aging Testing Function 5 3 1 Set up the parameters Step 1 Turn on Step 2 Press Shift 6 Aging Set to enter Ba...

Страница 28: ...tep 3 Press shift 3 to enter Battery Aging Test Function The lower right corner displays the current number of loops As shown below DICH Discharging CHAr Charging REST Rest Tips During the test you can press Shift 3 to suspend the test the data will be saved in the system Press Shift 3 to continue 5 3 3 List of test data Step 1 Turn on Step 2 Press Shift 9 to enter List of test data Step 3 Spin th...

Страница 29: ...29 Advanced Testing Function 5 3 4 Testing Process SHIFT 3 Aging Start DICH discharge CHAR charge DICH REST Loop Times Over END NO YES ...

Страница 30: ...0 100W 100 300W Resolution 0 001W 0 01W Accuracy 1 0 1 FS 1 0 1 FS Constant Resistance CR Range 0 1 99Ω 100 4KΩ Resolution 0 01Ω 1Ω Accuracy 1 0 3 FS 1 0 8 FS Voltage Measurements Range 0 18V 0 500V Resolution 0 001V 10mV Accuracy 0 05 0 02 FS 0 05 0 025 FS Current Measurement Range 0 3A 0 30A Resolution 0 0001A 0 001A Accuracy 0 1 0 1 FS 0 2 0 3 FS Short Circuit SHT Current 30A Resistance 40mΩ Ba...

Страница 31: ...ture and humidity range 18 28 80 RH or less Operating temperature and humidity range 10 40 10 90 RH Storage temperature and humidity range 0 50 10 90 RH Power Supply 110V 220V AC 48 5Hz 52 5Hz Fuse 1A Slow Blow Maximum rated power 15VA Weight 6kg net 6 3 Dimensions Instruments AT851 User s Guide ...