Specification
5
Specification
In Appendix A, you will learn the following content:
Technique Data
General Specifications
Appearance & Dimension
5.1
Technique Data
The following accuracy can be reached while in the circumstances:
Temperature: 23
℃
±5
℃
Humidity:
≤
65% R.H.
Zero value adjustment: Open & Short circuit zero-setting before testing.
Warm-up time: >60 Mins
Calibration Time: 12 Monthes
Average accuracy of testing level: 10%
Average accuracy of testing frequency: 0.02%
Accuracy of test terminal: Basic accuracy: 0.25%
C:
0.25% (1+ Cx/Cmax+ Cmin/Cx )(1+Dx )( 1+ks+kv+kf )
L:
0.25% (1+ Lx/Lmax+ Lmin/Lx )(1+1/Qx )( 1+ks+kv+kf )
Z:
0.25% (1+ Zx/Zmax+ Zmin/Zx )(1+ks+kv+kf )
R:
0.25%(1+ Rx/Rmax+ Rmin/Rx )(1+Qx)(1+ks+kv+kf )
D:
±
0.0020(1+ Zx/Zmax+ Zmin/Zx )(1+Dx+Dx
2
)(1+ks+kv+kf )
Q:
±
0.0025(1+ Zx/Zmax+ Zmin/Zx )(Qx+1/Qx )(1+ks+kv+kf )
Note
1.
L/C/R/Z are relative error, D/Q are absolute error
2.
Subscript “x” is measured value of the parameter, “max” is the maximum value, “min” is the
minimal value.
3.
ks is speed factor, kv voltage factor, kf is frequency factor
Following Max/Min value of the measurement parameters will affect the accuracy:
100Hz
120Hz
1kHz
10kHz
Cmax
800
µ
F
667
µ
F
80
µ
F
8
µ
F
Cmin
1500pF
1250pF
150pF
15pF
Lmax
1590H
1325H
159H
15.9H
Lmin
3.2mH
2.6mH
0.32mH
0.032mH
Zmax/ Rmax
1M
Ω
Zmin/ Rmin
1.59
Ω
Test speed error factor ks:
ks=0
Test level error factor: kv
Test level, instrument set parameters’ signal level V (rms), measure unit: mV
1Vrms
kv=0;
0.3Vrms
kv=1;
Test frequency error factor kf:
f = 100Hz,120Hz ,1kHz: kf=0
f = 10kHz kf=0.5