![Applent Instruments AT516 Скачать руководство пользователя страница 23](http://html.mh-extra.com/html/applent-instruments/at516/at516_user-manual_2954881023.webp)
23
<SETUP> Page
Comparator limit mode [MODE]
NOTE:
*1. This setting also can be set in the <MEAS DISPLAY> page.
5.3.1
Comparator
limit
mode
[MODE]
SCPI Command:
COMParator:MODE {ABS,PER,SEQ}
You can specify the primary parameter limit values in one of the following 2 ways:
Tolerance mode [%] [
]
In tolerance mode, the comparison limit values are based on the deviations from the
specified nominal value, which can be specified in the [NOM] field. You configure the
tolerance mode limit values as a deviation percentage (%) or absolute (
) parameter
value.
Sequential mode [SEQ]
In sequential mode, the comparison limit values are based on the absolute value of the
measurement. When you configure these limit values, you have to first define the
minimum value and then the maximum value.
Tolerance mode
Tolerance mode includes deviation percentage (%) or absolute (
).
Figure 5‐4
Tolerance mode
●
Includes the point
○
Excludes the point
Absolute value (
) = UNKNOWN value – nominal value
Deviation percentages (%) = Absolute value (
) / nominal value × 100%
Figure 5‐5
Example of sorting in tolerance mode
●
Includes the point
○
Excludes the point