22
AT8612 User's Guide
3.2
<NRM MEAS PAGE> Page
You can enter <NRM MEAS PAGE> page by pressing the [Meas] shortcut key.
<NRM MEAS PAGE> page mainly highlights the measured results, and display main parameter and sub
parameter.
In this page, you can finish the setting up of two basic functions, they include:
Basic Operating Mode – Set up four basic operating modes (CC/CV/CP/CR)
Input Value – Set up operating parameter
Note: measured data and sorting results are only valid in [Meas] Page.
Figure 3-2 <NRM MEAS PAGE> Page
3.2.1
Basic Operating mode
The basic four operation features of the Electronic Load are:
Constant current (CC) mode;
Constant voltage (CV) mode;
Constant power (CP) mode;
Constant resistance (CR) mode.
3.2.1.1
[CC] Constant current Mode
Under [CC] mode, the load will always consume constant current, no matter the input
voltage is changed or not.
If your measured power supply is constant voltage source, constant current mode is
recommended.
Figure 3-3 [CC] Mode
Setting
Current
Load
Current (I)
Load Input Voltage(V)