ATL-PV Test Leads  
Instruction 

 

Probe tip guard cap 

 

For CAT III or CAT IV environments, use the test leads 
with the probe tip guard cap fixed firmly. Without the 
probe tip guard cap, the test leads can be used in  
CAT II environment ONLY. 

 

For 1500V AC & 2000 V DC measurement, This test 
lead can only be used in the environment that is not 
connected to MAINS directly. 

 

Caution  

 

Make sure that test leads are firmly connected to the  
V-COM terminals of the correct instrument, and the 
instrument have to switch to PV mode. 

 

 

Warning  

When using test leads or probes, keep your fingers  

    behind the finger guards.

 

Use caution with voltages above 30 Vac rms, 42 Vac  

    peak, or 60 Vdc. These voltages pose a shock  
    hazard.

 

If the test lead is used in a manner not specified by  

    the manufacturer, the protection provided by the  
    equipment may be impaired.

 

To reduce the risk of fire or electric shock, do not use  

    this product around explosive gas or in damp  
    locations.

 

DO NOT USE the test leads when the internal white  

    insulation layer is exposed. 

DO NOT USE the test leads above maximum ratings  

    of CAT. environment, voltage and current, that are  
    indicated on the probe and the probe tip guard cap.

 

DO NOT USE the test leads without the probe tip  

    guard cap in CAT III and CAT IV environments. 

DO NOT USE the test leads to measure over     

   1000V that is connected to MAINS directly. 

Maintenance 

 

Do not attempt to repair this test lead set. It contains no 
user-serviceable parts. Repair or servicing should only  
be performed by qualified personnel. 

 

Cleaning

 

 

Clean the test lead with a water and mild detergent. DO 
NOT use abrasives or solvents and DO NOT IMMERSE  
in liquid. 
 

Specification 

 

Input Impedance: 

10MΩ 

Overvoltage Category: 

CAT II 1000V AC ,1500V DC 

                                        CAT III 1000V, CAT IV 600V. 

Pollution Degree 2 

 

Exposed probe tip length :  

19 mm to 4 mm (0.75 inch to 0.16 inch)  

Environmental ratings :  

-10°C to 45°C (-4°F to 113°F), 80% R.H. 

Altitude : 

2000 m (6,562 ft) 

Safety Standard : 

EN61010-031 

 
CAT          Application field

 

 

The circuits directly connected to Low-voltage 
installation. 

 

The building installation. 

 

The source of the Low-voltage installation. 

Finger Guard