Technical Data
SiOne Page
7
Measurement Technical Data
Version
Performance Parameters
Measurement Range
Measurement Accuracy (max.)
Caution
Si-B00…
Air pressure
Temperature
800…1100 hPa
14…140 ⁰F (-10…60 ⁰C)
±2 hPa
±0.9 ⁰F (±0.5 ⁰C)
Digital sensor, integrated
Si-CA0…
Carbon dioxide CO2
Max. 0…10000 ppm (0…1
%)
±40 ppm ±3% WO*
NDIR sensor, integrated
Si-CB0…
Carbon dioxide CO2
0.04…2 %
±0.02% ±3% WO*
NDIR sensor, integrated
Si-CE0…
Carbon dioxide CO2
0.04…5 %
±0.02% ±10% WO*
NDIR sensor, integrated
Si-D10…
Differential air pressure
Temperature
-100…+100 Pa
14…140 ⁰F (-10…60 ⁰C)
±0.1 Pa ±3% WO*
±3.6 ⁰F (±2.0 ⁰C)
Digital sensor, integrated
Si-D20…
Differential air pressure
Temperature
-250…+250 Pa
14…140 ⁰F (-10…60 ⁰C)
±0.1 Pa ±3% WO*
±3.6 ⁰F (±2.0 ⁰C)
Digital sensor, integrated
Si-D50…
Differential air pressure
Temperature
-500…+500 Pa
14…140 ⁰F (-10…60 ⁰C)
±0.1 Pa ±3% WO*
±3.6 ⁰F (±2.0 ⁰C)
Digital sensor, integrated
Si-H00…
Humidity
Temperature
0...100 %RH *
Max. -148...392 °F
(-100... 200 °C) *
±0.8 %RH *
±0.18 °F (±0.1 °C) *
External sensor (this version is
compatible with all available of
external APONE probes), with
dew point calculation function
Si-S00…
Humidity
Temperature
0...100 %RH
-4...140 °F (-20... 60 °C)
±2.0 %RH
±0.9 °F (±0.5 °C)
Digital sensor, integrated, with
dew point calculation function
Si-T00…
Temperature
14…176 ⁰F (-10…80 ⁰C)
±0.9 °F (±0.5 °C)
Digital sensor, integrated
Combi version - Q
Si-QHA…
Humidity
Temperature
Carbon dioxide CO2
0...100 %RH *
Max. -148...392 °F
(-100... 200 °C) *
Max. 0…10000 ppm (0…1
%)
±0.8 %RH *
±0.18 °F (±0.1 °C) *
±40 ppm ±3% WO*
2-in-1 model of the SiOne
(Si-H00… + Si-CA0…)
Si-QHB…
Humidity
Temperature
Air pressure
0...100 %RH *
Max. -148...392 °F
(-100... 200 °C) *
800…1100 hPa
±0.8 %RH *
±0.18 °F (±0.1 °C) *
±2 hPa
2-in-1 model of the SiOne
(Si-H00… + Si-B00…)