TROUBLESHOOTING AND CUSTOMER SUPPORT 

Verify Functionality 

Pressing the power button should activate the LCD and provide a real-time PPFD reading. Direct the sensor head 
toward a light source and verify the PPFD reading responds. Increase and decrease the distance from the sensor to 
the light source to verify that the reading changes proportionally (decreasing PPFD with increasing distance and 
increasing PPFD with decreasing distance). Blocking all radiation from the sensor should force the PPFD reading to 
zero. 

Battery Life 

When the meter is maintained properly the coin cell battery (CR2320) should last for many months, even after 
continuous use. The low battery indicator will appear in the upper left hand corner of the LCD display when the 
battery voltage drops below 2.8 V DC. The meter will still function correctly for some time, but once the battery is 
drained the pushbuttons will no longer respond and any logged measurements will be lost. 

Pressing the power button to turn off the meter will actually put it in sleep mode, where there is still a slight 
amount of current draw. This is necessary to maintain the logged measurements in memory. Therefore, it is 
recommended to remove the battery when storing the meter for many months at a time, in order to preserve 
battery life.  

Low-Battery Error after Battery Replacement 

A master reset will usually correct this error, please see the master reset section for details and cautions. If a 
master reset does not remove the low battery indicator, please double check that the voltage of your new battery 
is above 2.8 V, this is the threshold for the indicator to turn on. 

Master Reset 

If a meter ever becomes non-responsive or experiences anomalies, such as a low battery indicator even after 
replacing the old battery, a master reset can be performed that may correct the problem. Note that a master reset 
will erase all logged measurements from memory.  

Step 1:  press the power button so that the LCD display is activated. 

 

Step 2: Slide the battery out of the holder, which will cause the LCD display to fade out.  

Step 3: After a few seconds, slide the battery back into the holder. 

The LCD display will flash 

all segments and then show a revision number (e.g. “R1.0”). This indicates the master 

reset was performed and the display should return to normal. 

 

 

 

Содержание MQ-210X

Страница 1: ...NTS INC 721 WEST 1800 NORTH LOGAN UTAH 84321 USA TEL 435 792 4700 FAX 435 787 8268 WEB APOGEEINSTRUMENTS COM Copyright 2021 Apogee Instruments Inc OWNER S MANUAL UNDERWATER QUANTUM METER Model MQ 210X...

Страница 2: ...duction 4 Sensor Models 5 Specifications 6 Deployment and Installation 9 Battery Installation and Replacement 10 Operation and Measurement 11 Apogee AMS Software 14 Maintenance and Recalibration 16 Tr...

Страница 3: ...ectrical and electronic products with respect to the restriction of hazardous substances Please be advised that based on the information available to us from our raw material suppliers the products ma...

Страница 4: ...alibrated in air is used to make underwater measurements the sensor reads low This phenomenon is called the immersion effect and happens because the refractive index of water 1 33 is greater than air...

Страница 5: ...ee MQX series quantum meters covered in this manual are self contained and come complete with handheld meter and sensor Sensor model number and serial number are located on a label on the backside of...

Страница 6: ...m 2 s 1 Response Time Less than 1 ms Field of View 180 Spectral Range 370 to 650 nm wavelengths where response is greater than 50 of maximum see Spectral Response below Directional Cosine Response 5...

Страница 7: ...measurements were made at 10 nm increments across a wavelength range of 350 to 800 nm in a monochromator with an attached electric light source Measured spectral data from each quantum sensor were nor...

Страница 8: ...direct side by side comparison to the mean of seven reference SQ 500 quantum sensors Directional or cosine response is defined as the measurement error at a specific angle of radiation incidence Error...

Страница 9: ...or use with the MQ 210X to ensure the sensor is level when attached to a cross arm The bubble level in the plate makes leveling simple and accurate The AM 310 Sensor Wand accessory incorporates a moun...

Страница 10: ...m the meter circuit board NOTE The battery cradle can be damaged by using an incorrectly sized battery If the battery cradle is damaged the circuit board will need to be replaced To avoid this costly...

Страница 11: ...used to view and scroll through the logged measurements on the LCD display Logging To choose between manual or automatic logging push the mode button twice and use the up down buttons to make the appr...

Страница 12: ...struments com ac 100 communcation cable Spectral Errors Apogee SQ 100X series sensors can measure PPFD for sunlight and electric light with a single calibration factor However errors occur in various...

Страница 13: ...phenomenon is called the immersion effect and happens because the refractive index of water 1 33 is greater than air 1 00 The higher refractive index of water causes more light to be backscattered or...

Страница 14: ...s When the ApogeeAMS software is first opened it will show a blank screen until communication with the meter is established If you click Open Port it will say connection failed To establish communicat...

Страница 15: ...ls per day Click 30 Min Avg to see the meter s 99 30 minute averages To analyze the data click on File and Save As to save the data as a csv file Or you can highlight the numbers copy and paste them i...

Страница 16: ...app can be used to indicate the PPFD incident on a horizontal surface at any time of day at any location in the world It is most accurate when used near solar noon in spring and summer months where a...

Страница 17: ...s are available one for quantum sensors PPFD and one for pyranometers total shortwave radiation Clear Sky Calculator for quantum sensors Site data are input in blue cells in middle of page and an esti...

Страница 18: ...is still a slight amount of current draw This is necessary to maintain the logged measurements in memory Therefore it is recommended to remove the battery when storing the meter for many months at a t...

Страница 19: ...nal cable otherwise two splices would need to be made between the meter and sensor head See Apogee webpage for further details on how to extend sensor cable length http www apogeeinstruments com how t...

Страница 20: ...factory The warranty does not cover equipment that has been damaged due to the following conditions 1 Improper installation or abuse 2 Operation of the instrument outside of its specified operating r...

Страница 21: ...nship Apogee Instruments will repair or replace the items free of charge If it is determined that your product is not covered under warranty you will be informed and given an estimated repair replacem...

Отзывы: