background image

2

3

4

Oxygen Meter

This sensor is a galvanic cell-type oxygen sensor that 

measures oxygen gas (O

2

) in air. It has a lead anode, a gold 

cathode, an acid electrolyte, and a fluorine resin membrane. 

The current flow between the electrodes is proportional to 

the oxygen concentration being measure. An internal bridge 

resistor is used to provide an mV output linearly proportional 

to O

2

. Unlike polargraphic oxygen sensors, galvanic cells do not 

require a power supply. The hand-held meter used to measure 

the sensor uses a 3 V coin cell battery.

The mV output responds to the partial pressure of oxygen in 

air. The standard units for partial pressure are kPa. However, 

gas sensors that respond to partial pressures are typically 

calibrated to read out in mole fraction of the gas in air, or units 

of moles of oxygen per mole of air. These units can be directly 

converted to % O

2

 in air, or ppm O

2

 in air. The concentration 

of oxygen in our atmosphere is 20.95%, and this precise 

percentage has not changed for decades. It is also constant 

across changing temperatures or pressures. This allows for 

precise calibration of the instrument.

Being a galvanic cell type sensor, a small amount of oxygen 

is consumed in the reaction in order to produce the current 

flow and subsequent mV output. The oxygen consumption 

was measured to be 2.2 µmol O

2

 per day when the O

2

 

concentration was 20.95% (3240 mmol) at 23 C.

Humidity Changes

The graph below shows an example of humidity dependency. 

The sensor chemistry is not influenced by humidity, but its 

output decreases because O

2

 is displaced by water vapor 

molecules in the air. The effect of humidity is larger at warmer 

temperatures because there is more water vapor in the air. To 

eliminate humidity effects, simply recalibrate.

For use in high humidity, such as soil, remove 

the head and take the calibration measurement 

over water in a sealed container as shown at 

right.

Oxygen Meter

MO-200 Series

Effects on Output

For the most reliable measurements, the O

2

 sensor is designed 

to be used in a vertical position, as shown. A flow-through or 

diffusion head is not required for making measurements.

Influences from Various Gases

The sensor is unaffected by CO, CO

2

, NO, NO

2

, H

2

S, H

2

, and 

CH

4

. There is a small effect (approximately 1%) from NH

3

, HCI, 

and C

6

H

6

 (benzene). The sensor is sensitive to SO

2

 and can be 

damaged by O

3

.

Отзывы: