*IP55
The fan inside the TS-100 is extremely well sealed against water and dust, with an ingress protection (IP) rating of
55, called IP55. Higher numbers indicate better protection. Fans typically range from IP11 to IP55. The first digit in
the IP rating indicates protection against particle entry. A 5 indicates minimal dust entry with no interference with
fan operation. The second digit indicates water protection. A 5 indicates that water sprayed on the fan from any
direction shall have no harmful effects. Typical fans have particle and water protection ratings of 2 to 3, indicating
moderate particle and water protection, but no dust protection.
ST-110 (Precision Thermistor, not included w/ TS-100 or TS-120; included w/ TS-110 & TS-130)
TS-100/110/120/130
Difference among
Individual Replicate
Shields
Less than 0.1 C
Aspiration Rate
6 m s
-1
at full-speed; 3 m s
-1
at half-speed
Fan Input Voltage
Requirement
10.8 to 13.2 V DC
Fan Current Draw
80 mA at full-speed; 25 mA at half-speed
Fan Dust and Water
Protection
IP55*
Operating Temperature
-40 to 70 C
Dimensions
220 mm height, 270 mm diameter
Mass
840 g
Cables
5 m of shielded, twisted-pair wire for fan and air temperature sensors; TPR jacket (high water
resistance, high UV stability, flexibility in cold conditions); pigtail lead wires; stainless steel
(316), M8 connector located 25 cm from sensor head
ST-110
Measurement Range
-50 to 70 C
Measurement
Uncertainty
0.1 C (from 0 to 70 C); 0.15 C (from -35 to 0 C)
Measurement
Repeatability
less than 0.01 C
Non-stability
(Long-term Drift)
less than 0.02 C per year (when used in non-condensing environments where the annual
average temperature is less than 30 C; continuously high temperatures or continuously humid
environments increase drift rate)
Equilibration Time
5 s
Self-Heating
less than 0.01 C (typical, assuming pulsed excitation of 2.5 V DC); 0.08 C at 5 C (maximum,
assuming continuous input excitation of 2.5 V DC
Operating Environment
-50 to 70 C; 0 to 100 % relative humidity
Input Voltage
Requirement
2.5 V DC excitation (recommended, see (OPERATION AND MEASUREMENT section)
Output Voltage Range
0 to 2.5 VDC (assuming continuous input excitation of 2.5 V DC)
Current Drain
0.1 mA DC at 70 C (maximum, assuming continuous input excitation of 2.5 V DC)
Dimension
70 mm length; 2 mm diameter
Mass
60 g