![Apogee Instruments SIF-111 Скачать руководство пользователя страница 9](http://html.mh-extra.com/html/apogee-instruments/sif-111/sif-111_owners-manual_2954254009.webp)
Field of View
The field of view (FOV) is reported as the half-angle of the apex of the cone formed by the target surface (cone
base) and the detector (cone apex), as shown below, where the target is defined as a circle from which 98 % of the
radiation detected by the radiometer is emitted.
Sensor FOV, distance to target, and sensor mounting angle in relation to the target will determine target area.
Different mounting geometries (distance and angle combinations) produce different target shapes and areas, as
shown below.
CIRCULAR
HORIZONTAL
FOV Area (m²)
Half Angle
FOV Area (m²)
Type
Model
Half Angle
0 Deg
45 Deg
60 Deg
Type
Model
Horizontal
Vertical
0 Deg
45 Deg
60 Deg
Standard
SI-111
22°
2.069
7.627
45.211
Standard
SI-1H1
32°
13°
2.276
7.148
45.512
Narrow
SI-121
18°
1.337
4.461
18.865
Narrow
SI-4HR
16°
5°
0.403
1.15
3.341
Ultra-Narrow
SI-131
14°
0.787
2.447
8.544
-
A simple FOV calculator for determining target dimensions based on infrared radiometer model, mounting height,
and mounting angle, is available on the Apogee website: