IMPORTANT:
Transferring stored data requires removal of the bottom cover, which will result in the loss of all
current-day data.
1. Using a small Phillips-head screwdriver, turn the two screws
at the bottom until they turn freely (this should take roughly 5
turns).
Note: The screws do not need to be fully removed.
2. Remove the bottom cover. With the screws loosened, the
batteries should begin to push off the cover. If needed, you
can use the ¼-20 threaded mount with a matching bolt or
screw to pull off the cover. If the cover seems stuck, verify the
screws are completely unthreaded.
3. Connect a USB-C cable to the DLI meter and a computer.
Locate the meter (titled DLI METER) in file explorer as you
would a thumb drive. There will be three files in this drive:
ABOUT.TXT
This file contains calibration, firmware, and hardware information about your meter.
DLI.CSV
This four-column file contains the completed day number, days-ago label, DLI, and
photoperiod data up to 99 days.
MEAS.CSV
This two-column CSV contains a record number and 30-minute PPFD averages up to the
most recent 10 days.
Note: It can take up to 10 seconds for the DLI meter to connect to your computer.
4. Copy files to your computer.
5. Disconnect the USB-C cable.
6. Replace the bottom cover. Using a small Phillips screwdriver, replace the screws and tighten until snug.
Note: Do NOT over-tighten the screws.
Содержание DLI-400