![Apex Digital Sturdy SA-260MA-R Скачать руководство пользователя страница 35](http://html.mh-extra.com/html/apex-digital/sturdy-sa-260ma-r/sturdy-sa-260ma-r_instruction-manual_2952912035.webp)
32
6.6 Flash Program (optional, exclusive either to Re-Dry or Liquid function)
A. Before start Sterilization program please refer to
“6.2 Prepare Sterilization” section.
B. How to set the Flash program
Figure 46
C. Press
to start “Flash Program”. The corresponding sterilization time will be:
Sterilization Time
4 min.
Sterilization Temperature
134°C
Dry Time
1 min
D. Press the START button
, the setting temperature, duration, and dry time displayed two
times on the three-digit display, and then the sterilizer will automatically run through the
selected program. The current progress of the sterilization cycle is indicated by the
illuminated LED on the Process Status Indicator. (Figure 47)
Figure 47
134°C
(Room Temp.)°C
0 bar
(kgf/cm
2
)
Vacuum