7. Troubleshooting
If your Mini-Plus Compressor Nebulizer fails to function, consult the Troubleshooting guide below. If the
problem persists, consult your equipment provider.
Problem
Cause and Solution
Device
doesn’t
operate.
•
Check if plug is properly fit into an appropriate electrical outlet.
•
When device has been run continuously for over 30 minutes right
before using, an auto shut down may activate by built-in thermal
protector, cool down device for 30 minutes before next usage.
•
Check if fuse is blown and properly situated. (see “Maintenance”
section for details)
Weak
Nebulization
•
Check for proper electrical voltage.
•
Check tubing for blockage or air leakage at connection to Mini-Plus
compressor or nebulizer cup, replace as needed.
•
Check the nebulizer cup if it is properly assembled and not damaged. If
there is any damage, replace as needed.
•
Check if filter is too dirty, replace as needed.