15
English
English
IEC61000-4-11
70 % U
T
(30 % dip
in U
T
)for 25 cycles
<5 % U
T
(>95 % dip
in U
T
)for 5 sec
70 % U
T
(30 % dip
in U
T
) for 25 cycles
<5 % U
T
(>95 % dip
in U
T
) for 5 sec
continued operation during
power mains interruptions, it is
recommended that the device
be powered from an
uninterruptible power supply or
a battery.
Power frequency
(50/60Hz) magnetic field
IEC61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to the application of the test level
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Immunity Test
IEC60601 test level Compliance Electromagnetic
Environment-Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms150 kHz to 80
MHz outside ISM
bands
a
3 V/m 80 MHz to
2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF
communications equipment should
be used no closer to any part of this
device, including cables, than there
commended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
P
d
2
.
1
150kHz to 80MHz
P
d
2
.
1
150kHz to 80MHz
P
d
3
.
2
80 MHz to 2.5G MHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
b
Field strengths from fixed RF
transmitters, as determined by an