18
TECHNICAL SPECIFICATIONS
Channel
Dual, isolated between channels
Pulse amplitude
Adjustable 0-80mA peak into 500 Ω load each channel, constant current
Pulse Rate
1Hz-160Hz (adjustable), 1Hz/step (1-20Hz), 5Hz/step (20-160Hz)
Pulse Width
a. 50-260μs adjustable for TENS, 10μs/step
b. 250μs fixed for EMS
Software ramp up
feature
By changing mode, the output will reset to zero then ramp up to its original setting
intensity. It will protect users from a sudden surge
Patient Compliance
counter
Shows the treatment times in hours
Patient Lock System
Prevents the user from changing any fixed parameters set by the physician or li-
censed practitioner
Timer
15, 30, 60 minute and continuous mode selectable
LCD
TENS - Show modes, pulse rate, pulse width, timer and CH1/CH2 on LCD panel
EMS - Show modes, pulse rate, on time, off time, timer and CH1/CH2 on LCD panel.