![Apera Instruments WS200 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/apera-instruments/ws200/ws200_instruction-manual_2952598019.webp)
- 19 -
Chart (7-2)
Solution
Temperature compensation coefficient
NaCl salt solution
2.12%/℃
5%NaOH solution
1.72%/℃
Dilute ammonia solution
1.88%/℃
10% hydrochroric acid solution
1.32%/℃
5% sulfuric acid solution
0.96%/℃
7.4.5.The other parameters setting, please see the chart (7-3)
7.4.6.
Special attention: when the content of parameter setting is
consistent with content of factory default setting, conductivity
measuring interface will display as picture (6-1); when the
content of parameter setting is different with content of factory
default setting, conductivity measuring interface will display as
picture (6-2); the code icon will appear on the upper right of
LCD, when parameters with code icon to be set more than two,
LCD just displays one of the codes; user should enter the
parameter setting and carefully check the contents of setting or
restore the meter to factory default setting and set up the parameters again.
7.5
Parameter Setting:
7.5.1. Conductivity parameter setting (chart (7-3))
Chart (7-3)
Prompt Mark
Parameter Setting Items
Code
Parameter
P1
Standard solution series
selection
7.6
USA (84μS/cm, 1413μS/cm, 12.88
mS/cm, 111.9 mS/cm)
CH
(146.6μS/cm,
1408μS/cm,
12.85mS/cm, 111.3mS/cm)
P2
Electrode constant selection
0.1, 1, 10
P3
Benchmark temperature
selection
25℃ 20℃ 18℃
Picture (6-1)
Picture (6-2)