OVERVIEW 

 

Thank you for purchasing Apera Instruments TN400 Portable Turbidity Meter (hereafter referred to as 

the instrument). 

 

The instrument uses infrared light emitting diode (LED 860±30nm) as the light source and 90° 

scattering method, which is compliant with ISO 7027 and DIN EN 27027 standards for the 

determination of    turbidity for water solutions. TN400 allows you to measure the turbidity of an aqueous 

sample both in laboratory and in the field.   

 

The instrument is an outstanding combination of the most advanced electronic technology and software 

design. With built-in microprocessor chips, elegant design, easy to use, the meter has the following 

remarkable characteristics: 

 

 

Smart  functions,  such  as  automatic  calibration,  operating  instructions,  automatic  power-off,  low 

voltage display, etc. 

 

Large TFT color screen, with blue background in measuring mode, green background in calibration 

mode, and text and graphics in operational directive. 

 

Operation guidance and reminders in the process of calibration, measurement and parameter setting. 

 

Average  measurement  mode  automatically  takes  10  times  consecutive  readings,  calculate  and 

display the average. It improves the accuracy of the measurement, especially suitable for sample 

solutions with rapid settling and continuous measurement changes. 

 

The instrument meets IP67 water-resistant grade and is suitable for use in harsh environments.

 

 

 

Everything comes in a rugged carrying case, including accessories such as calibration solutions, test 

vials, and more, convenient for use and carry. 

 

 

Содержание TN400

Страница 1: ...1 TN400 Portable Turbidity Meter Instruction Manual APERA INSTRUMENTS LLC aperainst com...

Страница 2: ...Display Mode 6 3 4 Keypad 7 3 5 Power Supply 7 4 Meter Calibration 8 4 1 Preparation for Calibration 8 4 2 Calibration Procedure Take 0 NTU and 20 NTU as an example 9 4 3 Notes for Calibration 10 5 T...

Страница 3: ...characteristics Smart functions such as automatic calibration operating instructions automatic power off low voltage display etc Large TFT color screen with blue background in measuring mode green bac...

Страница 4: ...s AMCO Polymer or Formazin Solution 0 20 100 400 and 800 NTU FNU Detector Silicon photovoltaic Measurement Mode Normal measurement and average measurement mode Display TFT color screen Sample Vials 25...

Страница 5: ...vial holder Close the cover when measuring Dust proof plug Take off the plug when measuring Housing Sample vial holder Display Calibration vials or sample vials Keypad Positioning mark Align the mark...

Страница 6: ...0 NTU Carrying case TN400 Turbidimeter Microfiber cloth Screwdriver to remove battery cover Silicone oil 10 ml AA Alkaline battery 4 Sample vial 6 Instruction manual Diagram 2 Measurement mode Measure...

Страница 7: ...erform single shot measurement long press the key to perform continuous measurement In calibration mode press the key to confirm the calibration In calibration mode press the key to select standard so...

Страница 8: ...your supplier to replace See the detail in Section 7 Replacement Parts b Replace zero turbidity solution Replace with 0 0 NTU calibration solution open the vial cap pour out the original solution add...

Страница 9: ...the vial to make the solution even For polymer calibration solutions that have not been used for a long time slowly flip the vial twice and let it stand for 2 minutes Be careful not to shake the solu...

Страница 10: ...00 NTU calibration vial in the sample vial holder and press to calibrate After calibration is done press to confirm repeat the steps to calibrate 400 NTU and 800 NTU After calibration is done press to...

Страница 11: ...ation solution If the wrong solution is selected for calibration the measurement will be completely wrong To fix this problem simply re calibrate the meter selecting the correct calibration solutions...

Страница 12: ...s of continuous measurement displaying 1 data every 8 seconds and will display the average value at the end see diagram 6 The average measurement mode can be used for observing the stabilizing process...

Страница 13: ...d for measurement Please do not wash the sample vial holder as it may damage the optical structure 6 WARRANTY We warrant this instrument to be free from defects in material and workmanship and agree t...

Страница 14: ...L 1 bottle T200 1 20 100 400 800 NTU AMCO polymer Compatible with TN400 TN480 4 bottles T500 3 Sample vials with lid 25 60mm 1 6 6 pcs TN400 S3 Silicone oil 10mL 1 bottle APERA INSTRUMENTS LLC Address...

Отзывы: