background image

 

- 8 - 

        icon  displays  to  ensure  measuring  accuracy;  when            icon  displays,  the  battery  must  be 

replaced, otherwise the instrument can't work properly. To extend battery life, user can setup auto-

power off timer in Parameter Setup P6 according to usage requirements 

3.  Use the screwdriver to remove the four screws on the battery cover and open the battery cover. 

Insert 4 pieces of “AA” alkaline battery. Please make sure the polarity is correct. Put on the battery 

cover and screw tightly. 

 

METER CALIBRATION 

 

4.1 

Preparation for Calibration 

a) 

Calibration Point 

The instrument has 5 calibration points: 0 NTU, 20 NTU, 100 NTU, 400 NTU, and 800 NTU. Among 

them,  0  NTU  point  uses  AMCO  0.0  NTU  calibration  solution  or  laboratory  distilled  water,  and  the 

remaining 4 calibration points use AMCO polymer solutions. Note that the cap of the 0.0 NTU solution 

vial  can  be  opened.  Once  the  solution  is  invalidated,  users  can  replace  the  0.0  NTU  calibration 

solution or laboratory distilled water. The AMCO 0.0 NTU calibration solution can be purchased from 

the  supplier;  For  the  remaining  4  calibration  solutions,  their  vial  caps  are  sealed  and  cannot  be 

opened. Simply dispose the solutions after they are expired and buy new ones from your supplier to 

replace. See the detail in Section 7 Replacement Parts. 

b)  Replace zero turbidity solution 

 

Replace with 0.0 NTU calibration solution: open the vial cap, pour out the original solution, add 

1/2 distilled water, close the cap and shake the vial to rinse it and pour out the water. Repeat it 

3 times. Shake off the distilled water in the vial. Pour in new 0.0 NTU calibration solution and 

tighten the vial cap. 

 

Replace with laboratory distilled water: open the vial cap, pour out the original solution, add 1/2 

distilled water, close the cap and shake the vial to rinse it and pour out the water. Repeat it 3 

times. Shake off the distilled water in the vial. Pour in new laboratory distilled water and tighten 

the vial cap. 

 

The accuracy of 0.0 NTU calibration solution and laboratory distilled water is same. But 0.0 NTU 

calibration solution has 6 to 12 months of shelf-life, distilled water can be contaminated easily 

after air contact and lose the validity in a couple of days. 

AMCO 0.0 NTU solution is provided in factory configuration. For purchase details, please refer to 

Section 7 Replacement Parts.   

c)  Clean vial surface 

Apply a small drop of silicone oil on the surface of the vial and wipe it off with a micro-fiber cloth to 

evenly distribute the silicone oil on the surface so as to cover smudges or scratches, which helps 

light scattering. But please pay special attention to the following points: 

 

Содержание TN400

Страница 1: ...1 TN400 Portable Turbidity Meter Instruction Manual APERA INSTRUMENTS LLC aperainst com...

Страница 2: ...Display Mode 6 3 4 Keypad 7 3 5 Power Supply 7 4 Meter Calibration 8 4 1 Preparation for Calibration 8 4 2 Calibration Procedure Take 0 NTU and 20 NTU as an example 9 4 3 Notes for Calibration 10 5 T...

Страница 3: ...characteristics Smart functions such as automatic calibration operating instructions automatic power off low voltage display etc Large TFT color screen with blue background in measuring mode green bac...

Страница 4: ...s AMCO Polymer or Formazin Solution 0 20 100 400 and 800 NTU FNU Detector Silicon photovoltaic Measurement Mode Normal measurement and average measurement mode Display TFT color screen Sample Vials 25...

Страница 5: ...vial holder Close the cover when measuring Dust proof plug Take off the plug when measuring Housing Sample vial holder Display Calibration vials or sample vials Keypad Positioning mark Align the mark...

Страница 6: ...0 NTU Carrying case TN400 Turbidimeter Microfiber cloth Screwdriver to remove battery cover Silicone oil 10 ml AA Alkaline battery 4 Sample vial 6 Instruction manual Diagram 2 Measurement mode Measure...

Страница 7: ...erform single shot measurement long press the key to perform continuous measurement In calibration mode press the key to confirm the calibration In calibration mode press the key to select standard so...

Страница 8: ...your supplier to replace See the detail in Section 7 Replacement Parts b Replace zero turbidity solution Replace with 0 0 NTU calibration solution open the vial cap pour out the original solution add...

Страница 9: ...the vial to make the solution even For polymer calibration solutions that have not been used for a long time slowly flip the vial twice and let it stand for 2 minutes Be careful not to shake the solu...

Страница 10: ...00 NTU calibration vial in the sample vial holder and press to calibrate After calibration is done press to confirm repeat the steps to calibrate 400 NTU and 800 NTU After calibration is done press to...

Страница 11: ...ation solution If the wrong solution is selected for calibration the measurement will be completely wrong To fix this problem simply re calibrate the meter selecting the correct calibration solutions...

Страница 12: ...s of continuous measurement displaying 1 data every 8 seconds and will display the average value at the end see diagram 6 The average measurement mode can be used for observing the stabilizing process...

Страница 13: ...d for measurement Please do not wash the sample vial holder as it may damage the optical structure 6 WARRANTY We warrant this instrument to be free from defects in material and workmanship and agree t...

Страница 14: ...L 1 bottle T200 1 20 100 400 800 NTU AMCO polymer Compatible with TN400 TN480 4 bottles T500 3 Sample vials with lid 25 60mm 1 6 6 pcs TN400 S3 Silicone oil 10mL 1 bottle APERA INSTRUMENTS LLC Address...

Отзывы: