reaction of electrode.

 

e)  The  electrode’s  measuring  tip  should  be  soaked  in  the  storage  bottle  containing  3M  KCL  storage 

solution to keep the membrane hydrated and junction unblocked. When measuring, please unscrew 

the bottle cap, pull out the electrode and rinse it with deionized or distilled water. After using, please 

put the electrode back into the bottle and screw tight the cap. Clean the bottle and replace the storage 

solution  if  the  storage  solution  gets  turbid  or  mildewed.  The  electrode  should  never  be  soaked  in 

purified water or buffer solution for long.

 

f) 

NOTE: All the tests should be sample tests, meaning after measurement, the samples should be 

abandoned and not be used for its original purpose again.

 

g)  After 1-year of use, we recommend replacing the aged electrode to achieve the best accuracy.

 

Technical Specifications

 

2.1 

Instrument Technical Specs

 

 

Technical Specs

 

pH

 

Range

 

-2.00 to 19.99 pH

 

Resolution

 

0.01/0.1 pH

 

Accuracy

 

±0.01 pH ±1 digit

 

Temperature 

Compensation

 

0 to 100 ˚C (Auto or Manual)

 

mV

 

Range

 

-1999 mV - 0 - 1999 mV

 

Resolution

 

- 200 mV - 0 - 200 mV: 0.1 mV; Remaining: 1 mV

 

Accuracy

 

±0.1% F.S ±1 digit

 

Temp.

 

Range

 

0 to 100˚C (32 to 212˚F)

 

Resolution

 

0.1˚C

 

Accuracy

 

±0.5˚C ±1 digit

 

 

2.2 

Others

 

Data Storage

 

500 groups

 

Storage Content

 

Numbering, date, time, measurement, unit, temperature 

 

Output

 

USB

 

Power

 

AA Batteries * 3 / DC5V (USB port) 

 

IP Ranking

 

IP57 dust-proof and waterproof

 

Dimensions& Weight

 

Meter: (88×170×33 ) mm/ 313g  With case

(360×270×76) mm/ 1.6kg

 

Содержание SX811-PW

Страница 1: ...1 SX811 PW Portable pH Meter for Purified Water Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...elete 10 4 pH Measurement 11 4 1 Information regarding pH Calibration 11 4 2 pH Calibration use 3 point USA Standard as an example 13 4 3 Customized Calibration 14 4 4 Sample test 15 5 mV measurement...

Страница 3: ...n the box 23 9 Warranty 24 10 Appendix I Parameter setup and factory default setup 24 11 Appendix II Abbreviation glossary 25 12 Appendix III Self diagnosis information 25 13 Recommended pH Electrodes...

Страница 4: ...e data protection your data is safe even with a dead battery Advanced digital processing technology improvs meter responding time and accuracy 1 3 points automatic calibration calibration guide and se...

Страница 5: ...aining in the surface of the movable sleeve which can greatly affect the correct measurement 1 3 2 Usage and Maintenance a Prior to measurement remove the blue rubber plug to maintain proper air press...

Страница 6: ...measurement the samples should be abandoned and not be used for its original purpose again g After 1 year of use we recommend replacing the aged electrode to achieve the best accuracy 2 Technical Spe...

Страница 7: ...C and RCA 3 Instrument Description 3 1 LCD Display 1 Measurement mode icons 2 Measurement value 3 Timing storage icon When this icon appears the meter is in the automatic storage mode 4 Date and time...

Страница 8: ...n when this icon appears the meter connects the computer 17 Low battery icon when this icon appears please renew the battery 3 2 Keypad 3 2 1 Keypad operations Short press 1 5 seconds Long press 1 5 s...

Страница 9: ...enu In submenu press this key to enter parameter set up mode In parameter set up mode press this key to confirm parameters Long press In pH measurement mode press and hold this key to change the resol...

Страница 10: ...ue 3 4 2 Automatic timing storage Set the storage timing eg 3 minutes from parameter P4 1 icon appears on LCD and the meter enters into the timing storage mode When press key icon flashes and the firs...

Страница 11: ...t with clean tissue paper DO NOT rub or wipe the probe just use paper to dap off excess water Unplug the blue rubber plug insert the probe into the test sample gently stir the probe for a few seconds...

Страница 12: ...er should be calibrated before test every time For ordinary accuracy measurements 0 1pH once calibrated the meter can be used for about a week or longer In the following cases the meter must be re cal...

Страница 13: ...rate the meter End icon appears after calibration is done The 1st point calibration is finished In the meanwhile CAL2 will flash at the upper right corner and 4 00 pH 10 01 pH will flash alternately a...

Страница 14: ...y to calibrate the meter After calibration is done LCD at the top right shows blinking CAL2 icon indicating the meter enters into the 2nd point customer defined calibration 3 Rinse pH electrode in dis...

Страница 15: ...m parameter setup P1 6 please see Section 6 3 PU 2 icon displays at the right top of LCD please refer to Diagram 11 Note In parameter setup either PU 1 or PU 2 measurement mode can be selected but bot...

Страница 16: ...rinciple According to the pH isothermal measurement principle the closer the test sample s temperature is to the calibration solution s the higher the accuracy of the measurement This principle is rec...

Страница 17: ...erge the electrode in 0 1mol L dilute hydrochloric acid for 30 minutes then wash it in distilled water then submerge it in the soaking solution for 6 hours b For organic or lipid pollutant clean the p...

Страница 18: ...ess key to return to the measurement mode P1 1 Select pH standard solution series P1 2 Set calibration due date P1 3 Check date of last calibration P1 4 Set pH measurement of purified water P1 5 Set p...

Страница 19: ...ment of distilled water Off On 1 Press key Off flashes then press key On flashes when parameter flashes press key to confirm Off turn off temperature compensation On turn on temperature compensation 2...

Страница 20: ...acklight timing 1 2 3 On 1 When press key 1 flashes then press key to select blinking 2 3 On When the parameter flashes press key to confirm Select On to turn on the backlight the time unit is minute...

Страница 21: ...rm Date display Month Date 2 After confirming the parameter press key to enter in P4 8 mode or press key to return to the measurement mode P4 8 Adjust time 1 Press key Hour flashes then press key Minu...

Страница 22: ...olved oxygen are classified in the file Export press this key to export the stored value to Microsoft Excel file Exit press this key PC Link program exits from the computer interface 7 3 Load software...

Страница 23: ...computer pH and mV are sorted in the program 7 6 Storage during operation During operation press key to store or set timing storage The measuring information is downloaded to the computer through USB...

Страница 24: ...rvice for the products purchased by customers not the service life of the tester or probe 10 Appendix I Parameter setup and factory default setup Modes Prompts Parameter setting items Abbreviation Des...

Страница 25: ...4 5 Clear readings P4 6 P4 7 P4 8 Others China United States of America NIST Off On No Yes 12 Appendix III Self diagnosis information Icons Self diagnosis information Wrong pH buffer solution or the b...

Страница 26: ...ed Water Low ion concentration samples LabSen 803 LabSen 813 Soil LabSen 553 Solid or semi solid samples cheese rice fruit etc LabSen 753 Strong acid samples LabSen 831 MP500 temp probe required Stron...

Отзывы: