- 6 -
reaction of electrode.
e) The electrode’s measuring tip should be soaked in the storage bottle containing 3M KCL storage
solution to keep the membrane hydrated and junction unblocked. When measuring, please unscrew
the bottle cap, pull out the electrode and rinse it with deionized or distilled water. After using, please
put the electrode back into the bottle and screw tight the cap. Clean the bottle and replace the storage
solution if the storage solution gets turbid or mildewed. The electrode should never be soaked in
purified water or buffer solution for long.
f)
NOTE: All the tests should be sample tests, meaning after measurement, the samples should be
abandoned and not be used for its original purpose again.
g) After 1-year of use, we recommend replacing the aged electrode to achieve the best accuracy.
2
Technical Specifications
2.1
Instrument Technical Specs
Technical Specs
pH
Range
-2.00 to 19.99 pH
Resolution
0.01/0.1 pH
Accuracy
±0.01 pH ±1 digit
Temperature
Compensation
0 to 100 ˚C (Auto or Manual)
mV
Range
-1999 mV - 0 - 1999 mV
Resolution
- 200 mV - 0 - 200 mV: 0.1 mV; Remaining: 1 mV
Accuracy
±0.1% F.S ±1 digit
Temp.
Range
0 to 100˚C (32 to 212˚F)
Resolution
0.1˚C
Accuracy
±0.5˚C ±1 digit
2.2
Others
Data Storage
500 groups
Storage Content
Numbering, date, time, measurement, unit, temperature
Output
USB
Power
AA Batteries * 3 / DC5V (USB port)
IP Ranking
IP57 dust-proof and waterproof
Dimensions& Weight
Meter: (88×170×33 ) mm/ 313g With case
:
(360×270×76) mm/ 1.6kg