Table-11 Temperature compensation coefficient of special solutions
Solution
Temperature compensation coefficient
NaCl solution
2.12 %/°C
5% NaOH solution
1.72 %/°C
Dilute ammonia solution
1.88 %/°C
10% hydrochloric acid solution
1.32 %/°C
5% sulfuric acid solution
0.96 %/°C
8.8
Notes about conductivity parameter setting
When the content of conductivity parameter setting is different to the
factory default setting, the correspondant code icon will appear on the
upper right of LCD in measurement mode. When more than one
conductivity parameter is changed, only one code will be displayed. You
need to enter the parameter setting and check the settings content or
restore the meter to factory default setting and set the needed parameters again.