background image

 

3. 

Complete kit 

 

 

 
 
 

4. 

Preparation Before Use 

1) Pour certain amount of Storage solution (KCl) into the probe cap, soak the probe for 3 to 5 
minutes to activate the sensor (if frequently used, there is no need to soak). 

 

2) The soaking solution is 3M KCl solution, which comes with the test kit. Users can also make it 
by themselves by dissolving 25g KCl in 100 ml distilled water. 

 

5. 

ORP Calibration 

1)  Clean electrode in distilled water and dry it.     

 

2)  Pour certain amount of 222 mV ORP standard buffer solution into the calibration bottles 

(about half volume of the bottle). 

 

3)  Short press         to turn the meter on. Long press        to enter calibration mode. If you need to 

exit calibration mode, short press 

 
4)  Dip electrode in the standard buffer solution, stir gently, and allow it to  
stand still in the calibration solution until a stable reading is reached. When  

stable icon 

comes up, press          to adjust the calibration value 

according to the table below(also on the buffer bottle

’s label).  

 
 
 
 
 
 
 

Graph - 2 

 

Содержание ORP60

Страница 1: ...ORP60 Premium Pocket ORP Tester APERA INSTRUMENTS LLC www aperainst com v3 1...

Страница 2: ...In calibration mode press to change calibration value unidirectional 2 In mode setting short press to change parameter unidirectional 1 Long press to enter calibration mode 2 In calibration mode shor...

Страница 3: ...on 1 Clean electrode in distilled water and dry it 2 Pour certain amount of 222 mV ORP standard buffer solution into the calibration bottles about half volume of the bottle 3 Short press to turn the m...

Страница 4: ...How to fix is pressed before the measurement becomes stable smiley face comes up Wait till the smile icon appears and stays then short press 7 1 The cleaning and activation of ORP probes After long ti...

Страница 5: ...of 120 mV and 150 mV Preset low value 120 mV highest value 150 mV when the measured value 120mV or 150mV stable icon displays on LCD Red back lit heads up displays B Automatic lock P3 Select On to ac...

Страница 6: ...table measurement Reading value auto hold HOLD Self Diagnostic Information Er2 Low Voltage warning flickering reminder of battery replacement Three color backlight Blue Measurement Mode Green Calibrat...

Страница 7: ...MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to Transportation storage improper use failure to follow the product instructions or to perform any preventi...

Страница 8: ...APERA INSTRUMENTS LLC Address 6656 Busch Blvd Columbus Ohio 43229 Phone 1 614 285 3080 Email info aperainst com Website www aperainst com...

Отзывы: