- 7 - 

 

The Maintenance of ORP Probes 

After long-time use, the surface of the platinum pin can be contaminated and cause inaccuracy or slow response time. 

If that happened, the following methods can be adopted to clean and re-activate the sensor: 

a)  For contamination caused by inorganic substances, dip the probe into 0.1 mol/L dilute hydrochloric acid for 

30 minutes. Clean it with distilled water, soak it into the KCL storage solution for 6 hours, and then it will be 

ready for use again. 

b)  For contamination caused by organic substances and oiliness, clean the platinum surface with dishwashing 

liquid, and rinse 

with distilled water. And then the probe will be ready for use again after 6 hours’ soaking in 

the storage solution. 

c)  If the platinum surface is so severely contaminated that there forms an oxide film, toothpaste can be used 

to clean the platinum surface, and then wash it with distilled water. The probe will be ready for use again 

after 6 hours’ soaking in the storage solution. 

 

Paramet er Setting  

 

Table of Settings 

 

Parameter Setting 

1) 

When the meter is turned off, long press           

to enter parameter setting →short press          to switch P01-P02… 

→P5. Short Press          , parameter flashes → short press            to adjust parameter → short press            to confirm 

→Short press          to exit parameter setting and go back to measurement mode. 

2) 

Auto. Lock (P02)

 

– Users can set the auto lock time from 5 to 20 seconds. For example, if 10 seconds is set, when 

the measured value is stable for more than 10 seconds, the measured value will be automatically locked, and the 

HOLD icon will be displayed. Short press          to release the lock. When the setting is "Off" , the Auto. lock function 

is turned off, that is, the measured value can only be manually locked. Short press          to lock or unlock the 

measured value. The HOLD icon will be displayed when reading is locked. 

3) 

Auto. Backlight (P03)

 

  Users can set the automatic backlight time for 1 to 8 minutes. For example, if 3 minutes 

is set, the backlight will turn off automatically after 3 minutes; when the 

“Off” is set, the auto. backlight function will 

be turned off, and short press          to manually turn the backlight on or off. 

4) 

Auto. Power off (P04)

 

  The auto. power off time can be set to 10 to 20 minutes. For example, if 15 minutes is 

set, the meter will automatically shut down after 15 minutes if no operation; when 

“Off” is set, the auto. power off 

function will be turned off. Long press          to manually shut down the meter. 

5) 

ORP Back to Factory Default (P05) 

– Select “Yes” to restore instrument calibration to theoretical value. This 

function can be used when instrument does not work well in calibration or measurement. Calibrate and measure 

again after setting the instrument back to factory default.   

Symbol 

Parameter Setting Contents 

Content 

Factory Default 

P1 

Temperature Unit 

˚C – ˚F 

˚F 

P2 

Select automatic lock 

5-20 seconds 

– Off 

Off 

P3 

Automatic Backlight Off 

1-8 minutes

 

– Off 

P4 

Automatic Power Off 

10-20 minutes 

– Off 

10 

P5 

ORP restores back to factory default 

On -- Off 

Off 

Содержание ORP60-Z

Страница 1: ...1 SKU AI3717 ORP60 Z Smart ORP Redox Tester ORP Temp Instruction Manual APERA INSTRUMENTS LLC www aperainst com V1 03...

Страница 2: ...arranty 8 ATTENTION 1 You may find a few drops of water in the probe cap These water droplets are added to maintain the sensitivity of the pH sensor before the product leaves factory It does NOT mean...

Страница 3: ...dial display 3 Graph Mode digital display graph display 4 Table Mode digital display real time measurement and history display Calibration Press buttons to operate Operate on smartphone following gra...

Страница 4: ...n off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 2 In parameter setting short press to change parameter Uni directional 1 Long press to enter calibrat...

Страница 5: ...ed e g distilled deionized water for a long time which will make the electrode response slow If this happens soak the ORP electrode in 3M KCl solution for 3 5 hours and then re calibrate it before usi...

Страница 6: ...ound 20 C then adjust the calibration value to 227 mV Notes 1 Usually there s no need to calibrate for ORP testing When the tester has been used for a long time users can test it in 222 mV ORP standar...

Страница 7: ...For example if 10 seconds is set when the measured value is stable for more than 10 seconds the measured value will be automatically locked and the HOLD icon will be displayed Short press to release...

Страница 8: ...ery This limited warranty does NOT cover any damages due to accidental damage unauthorized repair normal wear and tear or external causes such as accidents abuse or other actions or events beyond our...

Отзывы: