How to use 

1.  Connect the electrode to the BNC connector of your pH meter. 

2.  Before measuring, twist off the storage bottle cap (see graph on the right), pull out the 

electrode and rinse it off with distilled or deionized water. 

3.  Unplug the blue rubber plug to maintain a smooth electrolyte flow. 

4.  Perform  at  least  a  two-point  calibration  before  measuring  after  connecting  the  new 

electrode to your pH meter. 

5.  Measurement suggestions for HF-containing solutions: 

5.1.  Use a PTFE container while testing solutions containing hydrofluoric acid. 

5.2.  The electrode can measure HF solution less than 0.1mol/L (pH

2.1) within 3 minutes each time. 

5.3.  When  measuring  HF  solution  around  0.5mol/L  (pH

1.8),  after  a  long  test,  if  the  membrane  is 

obviously dissolved, the service life will be shortened. Our recommendation is to avoid such tests or 

test faster, less than 2 minutes each time. 

5.4.  The electrode is not suitable to measure HF solution more than 0.5mol/L (pH

1.8), which will cause 

the electrode to perform at a poorer level or even be damaged. 

5.5.  After  a  while  of  usage,  potential  drift  and  slow  response  may  occur  to  the  electrode.  Please  re-

calibrate  after  soaking  it  in  3M  KCL  for  more  than  half  an  hour.  Stirring  is  recommended  while 

measuring to speed up stable measurement. 

6.  After using, put the electrode back into the storage bottle, twist on the bottle cap, and plug in the refilling 

hole. 

Maintenance 

1.  When not in use, the electrode should be soaked in the storage bottle containing 3M KCL soaking solution 

(SKU: AI1107)  to  keep  the  glass  membrane  and  junction  in  a  healthy  condition.  Clean  the  bottle  and 

replace the soaking solution if it gets contaminated. The electrode should never be stored in pure water 

such as deionized or distilled water. 

2.  The reference solution will run low as you use the electrode. Whenever the solution level falls to 1/2 height 

of the electrode, add 3M KCL solution (SKU: AI1107) to the refilling hole (unplug the blue rubber plug) 

using a syringe or pipette. 

3.  The electrode is only as accurate as it is clean. Always thoroughly rinse off the probe before and after each 

measurement with pure water in a container or with a wash bottle.   

4.  For tough contaminants, soak the electrode in Apera cleaning solution (AI1166) for 30 minutes. Then use 