2.3 

Prior to measurement, remove the rubber plug to maintain pressure of the reference solution, 
keep consistent flow rate of reference solution and stable potentials of junction. 

2.4 

After a period of usage, the reference solution will running low. Whenever the level falls to 1/2 
height of the electrode, add 3M KCL solution to the refilling hole by using syringe or pipette. 

2.5 

The  connector  of  the  electrode  should  be  kept  clean  and  dry.  If  being  contaminated,  please 
clean it with medical cotton and absolute alcohol and blow dry to prevent the short circuit of the 
electrode and slow reaction of electrode. 

2.6 

The electrode measuring tip should be soaked in the soaking bottle containing certain amount 
of storage solution to keep the membrane hydrated and junction unblocked. Clean the bottle 
and replace the storage solution if the storage solution gets turbid and mildewed. The electrode 
should never be sinked in pure water or buffer solution for long. 

2.7 

Please avoid measuring dehydrated medium like strong acid or alkaline solution, absolute ethyl 
alcohol and concentrated sulfric acid. In case of measuring such solution, please try to reduce 
the immersion time and clean it carefully after use. 

2.8 

After 1-year of use, we recommend replacing the electrode to ensure the best accuracy. 

 

Limited Warranty 

We warrant this electrode to be free from defects in material and workmanship and agree to repair 
or replace free of charge, at option of APERA INSTRUMENTS (Europe) GmbH any malfunctioned 
or damaged product attributable to responsibility of APERA INSTRUMENTS (Europe) GmbH for a 
period of SIX MONTHS from the delivery.   

 
This limited warranty does not cover any damages due to: 
Transportation,  storage, improper  use, failure  to follow  the  product  instructions  or  to  perform  any 
preventive maintenance, modifications, combination or use with any products, materials, processes, 
systems or other matter not provided or authorized in writing by us, unauthorized repair, normal wear 
and  tear,  or  external  causes  such  as  accidents,  abuse,  or  other  actions  or  events  beyond  our 
reasonable control. 

 
 
 
 
 
 
 
 
 
 

APERA INSTRUMENTS (Europe) GmbH 

Wilhelm-Muthmann-Str.15, 42329 Wuppertal Germany   

Phone: +49 202 51988998 

Website: 

www.aperainst.de

   

Email: [email protected] 

Отзывы: