2.3
Prior to measurement, remove the rubber plug to maintain pressure of the reference solution,
keep consistent flow rate of reference solution and stable potentials of junction.
2.4
After a period of usage, the reference solution will running low. Whenever the level falls to 1/2
height of the electrode, add 3M KCL solution to the refilling hole by using syringe or pipette.
2.5
The connector of the electrode should be kept clean and dry. If being contaminated, please
clean it with medical cotton and absolute alcohol and blow dry to prevent the short circuit of the
electrode and slow reaction of electrode.
2.6
The electrode measuring tip should be soaked in the soaking bottle containing certain amount
of storage solution to keep the membrane hydrated and junction unblocked. Clean the bottle
and replace the storage solution if the storage solution gets turbid and mildewed. The electrode
should never be sinked in pure water or buffer solution for long.
2.7
Please avoid measuring dehydrated medium like strong acid or alkaline solution, absolute ethyl
alcohol and concentrated sulfric acid. In case of measuring such solution, please try to reduce
the immersion time and clean it carefully after use.
2.8
After 1-year of use, we recommend replacing the electrode to ensure the best accuracy.
3
Limited Warranty
We warrant this electrode to be free from defects in material and workmanship and agree to repair
or replace free of charge, at option of APERA INSTRUMENTS (Europe) GmbH any malfunctioned
or damaged product attributable to responsibility of APERA INSTRUMENTS (Europe) GmbH for a
period of SIX MONTHS from the delivery.
This limited warranty does not cover any damages due to:
Transportation, storage, improper use, failure to follow the product instructions or to perform any
preventive maintenance, modifications, combination or use with any products, materials, processes,
systems or other matter not provided or authorized in writing by us, unauthorized repair, normal wear
and tear, or external causes such as accidents, abuse, or other actions or events beyond our
reasonable control.
APERA INSTRUMENTS (Europe) GmbH
Wilhelm-Muthmann-Str.15, 42329 Wuppertal Germany
Phone: +49 202 51988998
Email: [email protected]