Battery Replacement
①
Pull up the battery cap (might take some force).
②
Slide the battery cover along the OPEN arrow to open the cover.
③
Open the battery cover.
④
Insert the batteries
(ALL POSITIVE SIDES FACING UP)
.
⑤
Press down the battery cover and hold it.
⑥
Slide the battery cover along the LOCK arrow to lock the cover.
⑦
Close the battery cap. Make sure to push it all the way down. The tester’s waterproof
rating may be compromised if the battery cap is not tightly closed.
9
Probe Replacement
•
Screw off the probe ring, unplug the old probe; plug in the new
probe (make sure to align the connector's position properly), and
screw on the probe ring.
Probes don’t last forever. Every probe will eventually age and fail even
if you don't use it that often. The typical service life of GroStar probes
is 18-24 months depending on the frequency of usage and how well you keep it clean and
properly stored.
We recommend replacing your probe at least every 18 months to ensure the best accuracy.
The nice thing about GroStar pens is that you can just buy a replacement probe instead of a
whole new tester.
10 Notes
10.1 Avoid testing in high (>113°F/ 45°C) or low temperature (<41°F/ 5°C) solutions as it will
cause greater measurement error and damage to the probe. Test your samples and
perform calibration close to room temperature as much as possible.
10.2 Make sure the battery cap is completely closed with the O-ring. Otherwise, the
waterproof rating could be compromised.