Probe Cleaning 

8.1 

The tester is only as accurate as the probe is clean. Always thoroughly rinse off the 

probe before and after each measurement with clean water in a container or with a 

wash bottle. 

 

 

 

 

 

 

 

 

 

 

 

8.2 

For tough 

contaminants, remove the sensor shield, soak the probe in Apera’s cleaning 

solution or detergent water for 30 minutes. Then use a soft brush to remove the 

contaminants. Afterwards, soak the probe in 3M KCL soaking solution for 1 hour. Rinse 

it off, then re-calibrate the tester before using again. 

 

Probe Storage 

9.1 

Under regular usage (daily or weekly use), make sure there are several 

drops 3M KCL soaking solution in the probe cap, and tightly close the cap 

with the O-ring. 

 

 

 

 

9.2 

For long-term storage (you are not going to use the product for more than a 

month), add 3M KCL soaking solution to the Fill line in the cap and store 

the probe in it. Close the probe cap tightly with the O-ring. 

 

If you find white crystals inside or outside the probe cap, it is perfectly normal. It is 

the 3M KCL soaking solution that crystalizes over time by its nature. Just rinse them 

off and add in new soaking solution. This chemical is not poisonous nor dangerous, 

and the probe’s performance will not be affected at all. 

 

or 

Содержание 28APE190352/AI21040

Страница 1: ...GroStar GS4 Premium pH EC Pen Tester User Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...ically for the horticultural market Since 1991 Apera instruments has been dedicated to providing hi tech accurate lab grade instruments and sensors GroStar s intelligent design reduces the guesswork s...

Страница 3: ...which you should pour the soaking solution or sample solution For details of probe storage see Section 9 Premium pH EC Combo Probe Upgraded double junction blue pH sensor increased durability and eas...

Страница 4: ...KCL soaking solution for 5 minutes before use 2 5 If the tester hasn t been used for a long time over 1 month please soak the probe in the 3M KCL soaking solution for at least 1 hour then calibrate i...

Страница 5: ...up to the right of Calibrating pH 10 is usually not necessary unless your estimated target pH is greater than 8 0 pH 3 5 will disappear in 30 days after calibration reminding you to recalibrate the te...

Страница 6: ...ential air bubbles then let it stand 5 2 Hold until screen turns green The pen starts automatic calibration Wait for Good to show up in 10 15 seconds indicating the calibration is completed then the p...

Страница 7: ...of very high levels as delicate plants cuttings and seedlings can experience fertilizer burn if the conductivity is excessively high EC 500ppm and 700ppm are simply different units preferred by differ...

Страница 8: ...in 9 Probe Storage 9 1 Under regular usage daily or weekly use make sure there are several drops 3M KCL soaking solution in the probe cap and tightly close the cap with the O ring 9 2 For long term st...

Страница 9: ...ver Every probe will eventually age and fail even if you don t use it that often The typical service life of GroStar probes is 18 24 months depending on the frequency of usage and how well you keep it...

Страница 10: ...s cleaning solution or detergent water Clogged junction Clean the probe with Apera s cleaning solution then soak it in 3M KCL soaking solution overnight Aged probe Replace the probe Testing pH of low...

Страница 11: ...of operation Backlight White measurement Green calibration Red errors Reading hold Manual Calibration is successful but reading is not accurate Aged probe Replace the probe Air bubbles around the sens...

Страница 12: ...terproof rating IP67 Reading stabilization icon 15 What s in the box 16 Accessories pH buffer Kit pH4 7 250ml each item AI1113 pH buffer Kit pH4 7 10 250ml each item AI1115 2 77 EC Standard Calibratio...

Страница 13: ...his warranty does not apply to replacement of items subject to normal use wear and tear and expressly excludes Cosmetic damage such as stains scratches and dents Damage due to accident improper use ne...

Страница 14: ...ES DO NOT ALLOW EXCLUSION OR LIMITATION FOR INCIDENTIAL OR CONSEQUENTIAL DAMAGES OR LIMITATION ON HOW LONG AN IMPLIED WARRANTY LASTS SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT BE APPLICABLE IF ANY P...

Отзывы: