background image

 

6.2 

For tough 

contaminants, remove the sensor shield, soak the probe in Apera’s cleaning 

solution or detergent water for 30 minutes. Then use a soft brush to remove the 

contaminants. Afterwards, soak the probe in 3M KCL soaking solution for at least 1 

hour. Rinse it off, then re-calibrate the tester before using again. 

 

Probe Storage 

7.1 

Under regular usage (daily or weekly use), make sure there are several 

drops 3M KCL soaking solution in the probe cap, and tightly close the cap 

with the O-ring. 

 

 

 

7.2 

For long-term storage (you are not going to use the product for more than a 

month), add 3M KCL soaking solution to the Fill line in the cap and store the 

probe in it. Close the probe cap tightly with the O-ring. 

 

If you find white crystals inside or outside the probe cap, it is perfectly normal. It is 

the 3M KCL soaking solution that crystalizes over time by its nature. Just rinse them 

off and add in new soaking solution. This chemical is not poisonous nor dangerous, 

and the probe’s performance will not be affected at all. 

 

Battery Replacement 

 Pull up the battery cap (might take some force). 

 Slide the battery cover along the OPEN arrow to open the cover. 

 Open the battery cover. 

 Insert the batteries 

(ALL POSITIVE SIDES FACING UP)

 Press down the battery cover and hold it. 

 Slide the battery cover along the LOCK arrow to lock the cover. 

 

Close the battery cap. Make sure to push it all the way down. The tester’s waterproof 

rating may be compromised if the battery cap is not tightly closed. 

Содержание 28APE190349/AI11010

Страница 1: ...GroStar GS1 Premium pH Pen Tester User Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...rket Since 1991 Apera instruments has been dedicated to providing hi tech accurate lab grade instruments and sensors GroStar s intelligent design reduces the guesswork so you can easily manage your cr...

Страница 3: ...s the level to which you should pour the soaking solution or sample solution For details of probe storage see Section 7 Premium pH Probe Built with low resistance lithium glass membrane with double ju...

Страница 4: ...e in 3M KCL soaking solution for 5 minutes before use 2 5 If the tester hasn t been used for a long time over 1 month please soak the probe in the 3M KCL soaking solution for at least 1 hour then cali...

Страница 5: ...up to the right of Calibrating pH 10 is usually not necessary unless your estimated target pH is greater than 8 0 pH 3 5 will disappear in 30 days after calibration reminding you to recalibrate the te...

Страница 6: ...1 If necessary you can manually hold lock the reading by short pressing Press it again to cancel the hold 5 2 Long press to switch temp unit between F and C 5 3 The tester will automatically power off...

Страница 7: ...store the probe in it Close the probe cap tightly with the O ring If you find white crystals inside or outside the probe cap it is perfectly normal It is the 3M KCL soaking solution that crystalizes...

Страница 8: ...as it will cause greater measurement error and damage to the probe Test your samples and perform calibration close to room temperature as much as possible 10 4 Never test oily liquids 10 5 Make sure...

Страница 9: ...ully immersed in the solution at least 1 inch deep Air bubbles around the sensor Make a quick stir in the solution to remove air bubbles Probe is not properly connected or the pin connector is broken...

Страница 10: ...alkaline batteries Backlight White measurement Green calibration Red errors Reading hold Manual Warranty Two years for the instrument one year for the probe pH probe Low resistance lithium glass membr...

Страница 11: ...proof of purchase as defined above and determines the Product is defective Apera may in its sole discretion either a repair the Product with new or refurbished parts or b replace the Product with a n...

Страница 12: ...D HAS NOT MADE ANY EXPRESS OR IMPLIED WARRANTY OR OTHER REPRESENTATION REGARDING THE PRODUCT AND DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE TO THE EXTENT PERMITTED B...

Страница 13: ...12 APERA INSTRUMENTS Europe GmbH Wilhelm Muthmann Stra e 18 42329 Wuppertal Germany Contact info aperainst de Website www aperainst de Tel 49 202 51988998...

Отзывы: