Preparation Before First Use 

2.1 

Pull out the battery insulation slip, and take off the probe cap. 

 

 

 

 

 

2.2 

Rinse off the probe in clean water, then shake off excess water.

 

 

 

2.3 

Perform a 2-point calibration at pH 7 and pH 4. For calibration tutorial, refer to Section 

3. 

 

2.4 

If you find the probe cap is dried out, soak the probe in 3M KCL soaking solution for 5 

minutes before use. 

 

2.5 

If the tester hasn’t been used for a long time (over 1 month), please soak the probe in 

the 3M KCL soaking solution for at least 1 hour, then calibrate it before measurement 

begins 

 

pH Calibration 

3.1 

Power on the pen and remove the probe cap. 

Always calibrate 

7.00 pH first. 

3.2 

Rinse the probe with clean water and shake off excess water, 

then submerge it in the 7.00 pH standard buffer; 

shake the probe 

up and down in the solution for a few seconds

 and let it stand. 

 

Содержание 28APE190349/AI11010

Страница 1: ...GroStar GS1 Premium pH Pen Tester User Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...rket Since 1991 Apera instruments has been dedicated to providing hi tech accurate lab grade instruments and sensors GroStar s intelligent design reduces the guesswork so you can easily manage your cr...

Страница 3: ...s the level to which you should pour the soaking solution or sample solution For details of probe storage see Section 7 Premium pH Probe Built with low resistance lithium glass membrane with double ju...

Страница 4: ...e in 3M KCL soaking solution for 5 minutes before use 2 5 If the tester hasn t been used for a long time over 1 month please soak the probe in the 3M KCL soaking solution for at least 1 hour then cali...

Страница 5: ...up to the right of Calibrating pH 10 is usually not necessary unless your estimated target pH is greater than 8 0 pH 3 5 will disappear in 30 days after calibration reminding you to recalibrate the te...

Страница 6: ...1 If necessary you can manually hold lock the reading by short pressing Press it again to cancel the hold 5 2 Long press to switch temp unit between F and C 5 3 The tester will automatically power off...

Страница 7: ...store the probe in it Close the probe cap tightly with the O ring If you find white crystals inside or outside the probe cap it is perfectly normal It is the 3M KCL soaking solution that crystalizes...

Страница 8: ...as it will cause greater measurement error and damage to the probe Test your samples and perform calibration close to room temperature as much as possible 10 4 Never test oily liquids 10 5 Make sure...

Страница 9: ...ully immersed in the solution at least 1 inch deep Air bubbles around the sensor Make a quick stir in the solution to remove air bubbles Probe is not properly connected or the pin connector is broken...

Страница 10: ...alkaline batteries Backlight White measurement Green calibration Red errors Reading hold Manual Warranty Two years for the instrument one year for the probe pH probe Low resistance lithium glass membr...

Страница 11: ...proof of purchase as defined above and determines the Product is defective Apera may in its sole discretion either a repair the Product with new or refurbished parts or b replace the Product with a n...

Страница 12: ...D HAS NOT MADE ANY EXPRESS OR IMPLIED WARRANTY OR OTHER REPRESENTATION REGARDING THE PRODUCT AND DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE TO THE EXTENT PERMITTED B...

Страница 13: ...12 APERA INSTRUMENTS Europe GmbH Wilhelm Muthmann Stra e 18 42329 Wuppertal Germany Contact info aperainst de Website www aperainst de Tel 49 202 51988998...

Отзывы: