- 23 -
7.3
Conductivity meter calibration (take 1413μS/cm calibration as an
example)
1. Long press
key to enter calibration mode, as
shown in left graph. Rinse pH probe in purified water,
remove excess water by shaking or filter paper.
Press
key to confirm.
2. Submerge probe in 1413μS/cm solution. Stir the
solution briefly and allow it to stay in the buffer
solution until
appears and stays. Press
key to finish calibration.
3. Calibration completed, instrument goes back to
measurement mode.
⚫
For mulit-point calibration, please repeat above 1~3 step until all the calibration is done. The
meter can perform calibration in same calibration solution until the value displayed is stable
and repeatable.
⚫
To quit calibration mode, please
key, meter will return to measurement mode.
⚫
Press
key to switch measurement mode
Cond
→
TDS
→
Salt.
7.4
Customized calibration (take 10μS/cm standard solution as an example)
1. Select User in parameter setting clause 2.2, press
key to return to measurement mode.