TECHNICAL SPECIFICATIONS
2.1
Technical Parameters
Measuring Method
ISO 7027 and DIN EN 27027
compliant nephelometric method (90°)
Light Source
Infrared light emitting diode (860±30nm)
Measuring Range
0 – 1000 NTU, automatic range selection:
0.01 – 19.99 NTU
20.0 – 99.9 NTU
100 – 1000 NTU
Accuracy
≤ ±2 % of r stray light
Repeatability
≤ ± 1% of reading or 0.02 NTU (FTU), the greater of the two
Resolution
0.01 / 0.1 / 1 NTU (FNU)
Stray Light
≤0.02 NTU (FNU)
Calibration Standards
AMCO Polymer or Formazin Solution:
0, 20, 100, 400 and 800 NTU (FNU)
Detector
Silicon photovoltaic
Measurement Mode
Normal measurement and average measurement mode
Display
TFT color screen
Sample Vials
Φ25×60 mm, high borosilicate glass with lid
Sample Volume
18 ml
Power Supply
4× AA Alkaline Batteries
Operating Condition
Temperature: 0 to 50°C
(
32°F to 122°F
);
Relative humidity
:
0 – 90% at 30°C, 0 – 80% at 40°C, 0 – 70% at 50°C, no
condensation
Storage Condition
Instrument: -40 to 60°C (-40 to 140°F)
Calibration Solution: 5 to 30°C (41 to 86°F)
Instrument sealing
grade
IP67
Certificates
CE and RoHS
Limited Warranty
2 years
Dimension and Weight
Meter: (90×203×80) mm/385g
Test Kit: (310×295×110) mm/1.5kg