![Anyload OCSD Series Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/anyload/ocsd-series/ocsd-series_operation-manual_2948868019.webp)
no load and it is in stable status.
Press
【
ENTER
】
key
〖
LoAd
〗
Hang the span calibration weight. In
this example we will use a 3000kg
weight. Wait until STB signal
appearing in the screen
Press
【
ENTER
】
key
〖
05000
〗
It will display the full scale capacity
you set in parameter settings. The
first digit will be flashing. Wait until
“STB” signal is displaying on the
screen.
Press
【
ZERO
】
or
【
DATE
】
and
【
TARE
】
key
〖
03000
〗
We will change the value to 3000
since we will be using a 3000kg
calibration weight. Press
【
ZERO
】
to
change value, press
【
TARE
】
to move to next digit.
Press
【
MENU
】
or
【
MODE
】
key
〖
-------
〗
Confirm, store and run the span
calibration
〖
-oL-
〗
〖
End
〗
When End is displaying meaning
calibration is finished.
〖
3000
〗
Attention: The test load required in order the calibration will continue should
be more than 20% of its FS capacity. It is highly recommended to use a test
load of at least 80% of its FS capacity to get a better calibration result.
7.3.2 Three Point Calibration
If you want the scale to have high linearity result, calibrate the scale using the 3
point calibration procedure:
Turn on the dynamometer. Make sure it is displaying
〖
0
〗
even there is
tare weight on it.
Turn the indicator off. While the indicator is off, press and hold
【
MENU
】
or
【
MODE
】
key then press the ON/OFF key until it displays
〖
LOCAL
〗
. For P180,
you must use its USB and insert it to a PC to get a power.
18
|