background image

Product Information 

PA 5 – Battery Exchange 

www.anton-paar.com | Created by Kreisel, Gudrun 22.03.18 | H17IE006EN-A_ProdInfo_OBA1_BatteryExchange.docx 

page 2 of 2 

 

 

Figure 3: Push the plug apart with the screwdriver 

6.  After you opened the plug you can remove the flat battery and insert a new battery (mat. no. 106781). 
7.  Close the plug of the new battery. 
8.  Close the cover of the PA 5 manometer carefully. 

 

1.2  Synchronizing PA 5 with a Computer 

NOTICE 

After exchanging the battery, you have to synchronize the manometer time with the computer time. 
When no synchronization was carried out, subsequent tests cannot be read out from the PA 5, because the 

manometer time and date cannot be processed by the PetroNet software. 

 

1.  To synchronize the PA 5 time with your computer time open your PetroNet software. 
2.  Connect the PA 5 to the PetroNet software as described in the instruction manual. 
3.  Click on the button <Synchronize PA 5 time with PC> as shown in Figure 4. 

The PetroNet software synchronizes know the PA 5 and your computer. 

 

 

 

Figure 4: Synchronizing the PA 5 with your computer 

 
 
 
 
 
 
 
 
 
 
 
 