background image

Product Information 

PA 5 – Battery Exchange 

www.anton-paar.com | Created by Kreisel, Gudrun 22.03.18 | H17IE006EN-A_ProdInfo_OBA1_BatteryExchange.docx 

page 1 of 2 

1  Battery Exchange of PA 5 

You purchased an exchange battery (mat. no. 106781) for your PA 5. To carry out the battery exchange, please 

follow the instructions given in this product information. 
 

1.1  Battery Exchange 

1.  Before exchanging the battery, switch off the PA 5. 

 

NOTICE 

When removing the cover, never turn the cover completely because the cables inside the manometer may unplug 

or twist. 

 

2.  Open the PA 5 by carefully moving the cover back and forth.  
3.  In the opened PA 5 manometer is a small plug as shown in Figure 1. The plug connects the battery to the 

PA 5. 

 

 

 

Figure 1: Plug for the battery 

 

4.  To remove the battery, disconnect the battery first. 

To disconnect the battery open the plug soldered to the battery. Use the screwdriver supplied with the PA 5 

as shown in Figure 2. 

 

 

 

Figure 2: Screwdriver to open the plug soldered to the battery 

 

5.  To open the plug push the tip of the screwdriver into the central flange and push the two parts of the plug 

apart as shown in Figure 3. 

 

Plug 

Отзывы: