Lovis 2000 M/ME Low Volume – SOP 

Standard Operating Procedure for Low Sample Volume

Hollow plug

1.   Put the hollow plug into some cleaning liquid. 

2.   Fill a clean syringe (5) with a hollow needle (6) with cleaning liquid. Insert the 

hollow needle into the plug (7) and rinse the inside of the hollow plug.

Suitable diameters: Hollow plug 1.59 or 1.8 >> Hollow needle 0.7 (gauge 22)

 Hollow plug 2.5

>> Hollow needle 1.2 (gauge 18)

3.   Take the hollow plug and wipe it with tissue paper.

4.   Blow the inside dry using a hollow needle and the built-in air pump. 

See how to dry the short capillary, point 5.). 

3

Installing (Adjusted) Lovis Measuring Systems 

(Manager and Administrator only)

1.   Tap "Menu > Setup > Measuring System Settings > Lovis Module > Capillary Administration".

2.   Tap <

Add

>. There are two possibilities:

Select  "

Import Capillary Configuration

" to import 

factory adjustment data 

from the 

USB flash drive. You can also import user-created data of an existing capillary/ball 
combination, exported from another Lovis beforehand. Tap <

OK

>.

-

The available "Drive" is displayed. Select a capillary/ball combination or a list of 
combinations in the field "Target". Tap <

Next

>. 

-

If there is a list, select one of the combinations.

-

You can assign a new name to the combination.

Select  "

Create New Capillary

" if no capillary/ball combination data is available. Tap <

OK

>.

-

Choose ball type, capillary diameter with extension "

short

" and enter the serial number. 

Name the capillary/ball combination. Activate the checkbox to use this combination 
unadjusted to measure only runtimes or adjust the combination (see chapter 6, p. 5).

4

Filling the Short Capillaries

1.   Select a suitable capillary for the liquid’s viscosity:

2.   Take a clean microliter glass syringe and a clean hollow needle with Luer fitting. Fix the hollow 

needle to the syringe tip.

TIP:

Clean the microliter glass syringe with a suitable cleaning liquid for your sample before using it for 
the first time. If the glass syringe is not dry after cleaning it, remove traces of the cleaning liquid 
by drawing in approx. 10 µL of sample and ejecting it into a waste vessel.

Capillary measuring core ranges

a

 (dynamic viscosity) / from... to...

a.

Allows to measure the viscosity at all angles from 15° to 80°. 

1.59

1 mPa.s

20 mPa.s

1.8

13 mPa.s

300 mPa.s

2.5

70 mPa.s

1700 mPa.s

5

6

7

Содержание Lovis 2000 M/ME

Страница 1: ...lary dry with the built in air pump Plug a clean hollow needle 1 2 gauge 18 onto the Adapter Luer cone 13 of the air hose 2 Capillary 6 Hollow needle 13 Adapter Luer cone of air hose Insert the hollow needle into the capillary as far as possible Start the air pump with the button on the touch screen The drying time depends on the cleaning liquid on the temperature and on the ambient air humidity D...

Страница 2: ...r Lovis beforehand Tap OK The available Drive is displayed Select a capillary ball combination or a list of combinations in the field Target Tap Next If there is a list select one of the combinations You can assign a new name to the combination Select Create New Capillary if no capillary ball combination data is available Tap OK Choose ball type capillary diameter with extension short and enter th...

Страница 3: ...d 7 Slowly fill the sample by pushing the plunger downwards 8 While the liquid rises draw the hollow needle upwards Make sure it remains below the liquid level and below any air bubbles 9 Fill sample until the liquid reaches the tip of the capillary 10 Check the entire filling The capillary must be filled without any bubbles or particles If detecting a bubble see arrow wait till it rises to the to...

Страница 4: ...hem with the hollow needle 12 Check the filling again The capillary must be filled without any bubbles or particles 13 Select a suitable clean hollow plug 7 for the capillary 2 and close it Press the hollow plug firmly into the capillary Make sure not to insert air bubbles with the plug Turn the hollow plug to insert it deeper into the capillary If necessary remove a little sample with the hollow ...

Страница 5: ... filling of the capillary for bubbles and check that the Lovis stands in a stable position Refill the capillary and tap Improve Lovis suggests Apply if the Level Adjustment correction is 0 5 The Level Adjustment is valid and stored after tapping Apply 6 Adjusting Lovis Measuring Systems Manager and Administrator only TIP If you have ordered a factory adjustment import the adjustment from the USB f...

Страница 6: ...ion between adjustment and verification default 0 5 7 Fill the capillary once again with your standard liquid Do NOT change the ball Follow the instructions on the display When the verification is finished the runtime deviations for each adjusted angle are displayed TIP After a verification the runtime deviations to the values of the preceding original adjustment are displayed Typically these valu...

Страница 7: ...surement is successfully completed if the progress bar turns green and the message above reads Finished The Lovis Condition is valid Menu Item Lovis Module Parameters Default Value Temperature Settings Temperature Measuring Temperature 20 C Temperature Equilibration This is the time Lovis waits for the sample to reach the measuring temperature after starting measurement and after the capillary blo...

Страница 8: ...mple 7 Move the plunger up and down rapidly to improve the cleaning effect Draw the cleaning liquid back into the syringe Empty the syringe into a waste vessel 8 Repeat the two above steps two to three times until no traces of sample are left in the capillary If the ball is still inside let it roll out and rinse the capillary without the ball once 9 If necessary repeat the cleaning procedure with ...

Страница 9: ...r only as far out as the µL scale Do not use strongalkaline cleaning liquids that will corrode the glass 1 Empty the syringe into a waste vessel 2 Draw in some cleaning liquid through the hollow needle and move the plunger up and down repeatedly If necessary wait a bit for the cleaning liquid to mix with the sample Empty the syringe into a waste vessel 3 Repeat the above step two to three times un...

Отзывы: