Anritsu VectorStar MN4690B Series Скачать руководство пользователя страница 2

QSG-2

PN: 10410-00290  Rev. A

MN4690A QSG

2.

Connecting the Front Panel Cabling

The illustration below shows the front panel connections between the Test Set and VNA.

1. 

Place the VNA on top of the test set as shown above.

2. 

Disconnect the front panel RF cable loops from VNA ports 

b1

 and 

b2

 and any cables that are connected 

from VNA 

Port 1

 and 

Port 2

.

3. 

Using the four provided RF (K or V) male-to-male cables, or other semi-rigid or phase-stable male-to-male 

RF cables:

a.

Connect the Test Set 

To VNA b1

 output port to the far left VNA  

b1

 input port.

b.

Connect the Test Set 

To VNA b2

 output port to the far right VNA 

b2

 input port.

c.

Connect the Test Set 

To VNA Port 1 Source OUTPUT

 port to the VNA 

Port 1 Source

 output port.

d.

Connect the Test Set 

To VNA Port 2 Source OUTPUT

 port to the VNA 

Port 2 Source

 output port.

Figure 2.

Front Panel Test Set Connections

Power

Port 1

Port 2

Port 3

Port 4

To VNA

b1

To VNA

b2

+23 dBm max
+40 VDC max

+23 dBm max
+40 VDC max

+27 dBm MAX    AVOID STATIC DISCHARGE

ALL PORTS: CAUTION

Standby Operate

Frequency

Power

Sweep

Avg

Response

Display

Scale

Marker

Channel

Trace

Measurement

Calibration

Application

File

System

Help

Preset

To VNA

Port 1 Source

OUTPUT

To VNA

Port 2 Source

OUTPUT

1

3

2

2

3

3

3