3-6 Directivity and Test Port Match Verification
Chapter 3 — Performance Verification
3-10
PN: 10410-00730 Rev. D
MN469xC Series Multiport Test Set MM
37.
Change all DUT Connector settings to
K-Conn(F)
for MN4694C or
V-Conn(F)
for MN4697C in the One
Port Cal Setup dialog box. Click the
OK
Figure 3-3.
One Port Cal Setup Dialog Box