Chapter 3 — Performance Verification
3-5 S-Parameters Measurements Verification
MN469xC Series Multiport Test Set MM
PN: 10410-00730 Rev. D
3-5
The quality of the verification result is dependent on the degree of care taken by the user in maintaining,
calibrating, and using the system. The most critical factors are:
•
The stability and quality of the devices in the calibration kit and the verification kit.
•
The condition of the VNA test port connectors and test port cables.
•
The pin depths of all connectors and the proper torquing of connections. These same factors also affect
the VNA measurement quality.
Consult the reference manual supplied with Anritsu Calibration Kits and Verification Kits for proper use, care,
and maintenance of the devices contained in these kits.
Equipment Required
•
Personal Computer with Microsoft Windows Operating System
•
National Instruments GPIB interface
•
GPIB interface cable
•
Anritsu Calibration Kit (refer to
Table 3-1, “Required Calibration and Verification Equipment
•
Anritsu Verification Kit (refer to
•
Anritsu Test Port Cables (refer to
)
Table 3-1.
Required Calibration and Verification Equipment
Model
Calibration Kit
Verification Kit
Test Port Cable
MS4644A/B+MN4694C
MS4644A/B+MN4694C
3652A
3668-1
3670K50-2 (Qty 2)
MS4645A/B+MN4697C
MS4647A/B+MN4697C
3654D
3669B-1
3670V50A-2 (Qty 2)
Note
The use of non-Anritsu calibration kits or verification kits is not supported.