Step 10.
Press the
Measure
soft key and follow the on-screen prompts to complete the
measurement.
NOTE: Access to the transmitter is required to complete this procedure as the
transmitted carrier must be turned off for the second portion of the measure-
ment.
Step 11.
After the measurement is complete, a message box will give results for the three
different signal types. Some measurement results may show as N/A, and this is
to be expected. Press the
ENTER
key to continue. All of the information in the
message box will still be available.
Step 12.
Press the appropriate soft key for the signal type being measured to view the
measurement results. The results are shown at the bottom of the screen.
Step 13.
Press the
Save C/I Meas
soft key to save both traces. Select the alphanumeric
characters for the trace name and press the
ENTER
key to save the traces. The
Site Master will save both the carrier and interferer traces with the same name,
and automatically append a “-C” to the saved carrier trace, and a “-I” to the
saved interference trace.
Step 14.
Press the
C/I Done
soft key to return to normal sweeping mode.
Step 15.
To recall the carrier and interferer traces, select the
RECALL DISPLAY
key
and use the Up/Down arrow key to select either the -C or -I trace. Both traces
will be displayed, with the selected trace data shown at the bottom of the dis-
play. With the standard monochrome display, the selected trace is displayed as a
black trace and the other as a gray trace. With the Option 3 color display, the se-
lected trace is displayed as a yellow trace, and the other as a magenta trace.
(Some adjustment of the display contrast (see page 2-7) may be required for op-
timal viewing.)
Step 16.
To review the specific data for the measurement, select the
MEAS/DISP
key
followed by the
Measure
soft key, the
More
soft key and the
C/I
soft key. Select
either the
C Data
,
I Data
, or
C/I Data
soft keys to view the corresponding data at
the bottom of the display.
Step 17.
Press the
ESCAPE
key to return to normal sweep mode.
5-21/5-22
Chapter 5 Spectrum Analyzer Measurements
Figure 5-14.
Recalled C/I Trace