Calibration
3-4 Station Asset Utility
MG362x1A MM
PN: 10370-10386 Rev. B
3-9
Set the Communication Parameters
Set the ATE communication parameters as shown in
Test Procedure
Set the communication parameters as follows:
1.
Set the Interface as TcpipSocket.
2.
Enter the MG362x1A IP Address.
3.
Enter the Port as 9001.
4.
Press the Add DUT button.
When all the test equipment assets have been entered and the communication parameters are set, the
automated testing is ready to perform testing.
1. TcpipSocket
2. IP Address of MG362x1A
3. Port set to 9001
4. Add DUT Button
Figure 3-4.
Set Communication Parameters
2
1
3
4