MA24xxA/B/D, MA2400xA OM
PN: 10585-00004 Rev. T
A-1
Appendix A — Test Records
A-1
Introduction
This appendix provides tables for recording the results of the
performance verification tests (
). They provide the means for
maintaining an accurate and complete record of instrument
performance.
A-2
Measurement Uncertainty
All test records are provided with a measurement uncertainty, which
consists of the type-B
1
components. The error contributions are
measurement method, test equipment, standards, and other correction
factors (for example, calibration factors and mismatch error) per the
prescribed test procedure. The statement(s) of compliance with
specification
2
is based on a 95% coverage probability for the expanded
uncertainty of the measurement results on which the decision of
compliance is based. For some of the measured values it is not possible
to make a statement of compliance with specification
2
. Other values of
coverage probability for the expanded uncertainty may be reported,
where practicable.
A-3
Test Records
We recommend that you make a copy of the test record pages each time
a test procedure is performed. By dating each test record copy, a
detailed history of the instrument’s performance can be accumulated.
1.
BIPM JCGM 100:2008 Evaluation of measurement data–Guide to the
expression of uncertainty in measurement.
2.
LAC–G8:03/2009: Guidelines on the Reporting of Compliance with
Specification.