7.2 Performance Test for Error Detector
7-19
7
Performance T
est
7.2.4
Maximum input level and patterns
(1) Specification
800 mVp-p
BER 10
–
12
or less under the following conditions:
Bit
rate:
When MP2110A-093 is not installed: 24.3 to 28.2 Gbit/s
When MP2110A-093 is installed:
9.5 to 28.2 Gbit/s
Pattern:
PRBS31, PRBS23, PRBS15, PRBS9, PRBS7
Mark
ratio: 50%
Single-end
Back-to-back
connection
(2) Setup
Figure 7.2.4-1 ED1 Data In Pattern Test Connection Diagram
Figure 7.2.4-2 ED1
Data In Pattern Test Connection Diagram
(3) Procedure
1. Connect coaxial terminators to the
PPG1
Data
Out
and
ED1
Data
In
connectors. (Refer to Figure 7.2.4-1.)
2. Connect
the
PPG1 Data Out
and
ED1 Data In
connectors using a
coaxial cable.
Coaxial Terminator
Coaxial Terminator
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...