Chapter 6 How to Operate Sampling Scope
6-62
6.3.4
Performing Self Test
Perform the self test in the following cases:
●
When the error message is displayed after level calibration referring to
●
When the noise voltage is over the specified value shown in Appendix A
when the signal is not input
●
When the waveform is not displayed, the displayed waveform
amplitude and bit cycle is different from the expected value, and the
MP2110A operation and measurement results are abnormal.
The application self test runs following items.
●
Power
If the test result is normal, “passed” is displayed, on the other hand, if the
test result is abnormal, “failed” is displayed.
The self test procedures are as follows.
1. Check that no signals are input to the front-panel
Ch A In
,
Ch B In
,
Trigger Clk In
connectors. If there are
Ch C
and
Ch D
, check that no
signals are input to these connectors.
2. Click
Setup to open the Setup dialog box.
3. Click
the
Utilities tab.
4. Click
Application Test. The self test dialog box is displayed while the
self test is running.
5. After the self test is completed, the result is displayed.
When the power supply test result is failed:
1. When the power voltage is displayed, record the value.
2. If the power supply test result shows “failed”, contact an Anritsu
Service and Sales office.
Содержание BERTWave Series
Страница 26: ...VI...
Страница 74: ...Chapter 1 Outline 1 48...
Страница 166: ...Chapter 4 Screen Operation 4 24...
Страница 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Страница 237: ...6 2 Explanation of Windows 6 43 6 How to Operate Sampling Scope Figure 6 2 4 9 Measure Dialog Box Equalizer Tab...
Страница 239: ...6 2 Explanation of Windows 6 45 6 How to Operate Sampling Scope Figure 6 2 4 10 Measure Dialog Box Mask Test Tab...
Страница 246: ...Chapter 6 How to Operate Sampling Scope 6 52 Figure 6 2 6 2 Time Dialog Box CRU Tab When MP2110A 054 is installed...
Страница 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Страница 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Страница 377: ...8 6 Adding an Option License 8 11 8 Maintenance 10 Click System Menu System Information and confirm the added option...
Страница 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Страница 432: ...Appendix A Specifications A 36...
Страница 457: ...D 3 Sampling Oscilloscope D 11 Appendix D 26G 53G CRU MP2110A 055 Table D 3 7 CRU Output Sensitivity Pass Fail...
Страница 458: ...Appendix D Performance Test Record Form D 12...
Страница 466: ...Index Index 6...