OPERATIONAL TESTS
LOG MAGNITUDE DYNAMIC ACCURACY TEST
37XXXD MM
4-19/4-20
Figure 4-2.
Log Magnitude/Dynamic Accuracy Test Result
37XXXD Log Magnitude Dynamic Accuracy Test
Model:______________________
Serial Number:____________________
Date:________
Operator:____________________
Step Atten. ID:_____________________
--A--
Attenuation
Device Nominal
Value
--B--
Measured
Value (Avg)
--C--
Attenuation
Device
Characterized
(or Traceable)
Value
--D--
Composite
Uncertainty
--E--
Col. C + Col. D
(Upper Limit)
--F--
Col. C - Col. D
(Lower Limit)
Pass or Fail
–10 dB
0.107
–20 dB
0.105
–30 dB
0.106
–40 dB
0.105
–50 dB
0.156
–60 dB
0.210
–70 dB
0.330
–80 dB
0.597
Table 4-1.
Log Magnitude/Dynamic Accuracy Test
Содержание 37 D Series
Страница 3: ......
Страница 4: ......
Страница 12: ......
Страница 14: ...1 2 37XXXD MM Figure 1 1 Series 37XXXD Vector Network Analyzer System...
Страница 26: ......
Страница 35: ...REPLACEABLE PARTS PARTS ORDERING INFORMATION 37XXXD MM 2 9 2 10 Figure 2 2 Signal Source Parts Location Diagram...
Страница 36: ......
Страница 38: ......
Страница 62: ......
Страница 80: ......
Страница 82: ......
Страница 90: ......
Страница 104: ......
Страница 122: ......
Страница 143: ...REMOVE AND REPLACE SIGNAL SOURCE MODULES 37XXXD MM 8 23 Figure 8 11 Signal Source Module Assemblies Removal Details...
Страница 148: ......
Страница 152: ......
Страница 154: ......
Страница 156: ......
Страница 158: ......
Страница 160: ...D 4 37XXXD MM ME7808 BLOCK DIAGRAM ME7808 BROADBAND SYSTEM MAINTENANCE Figure D 1 ME7808B C Block Diagram...
Страница 184: ......
Страница 190: ......