10-5
PHYSICAL SETUP
See Figure 10-2 for an illustration of the connections necessary to per-
form an automatic calibration using the AutoCal module. Note that
the connection is very simple. Different power cable connectors are
used with the mechanical modules and the electronic modules to pre-
vent connecting the wrong power supply in error.
There is no on-off switch. When power is connected to the AutoCal
module, the LED labeled POWER should come on immediately. The
second LED, labeled OPERATE, should come on in about five minutes,
after the internal temperature control oven has stabilized. (Internal
temperature is held within a 5
o
C window.)
PHYSICAL SETUP
AUTOCAL
10-6
37XXXC OM
7
8
9
4
5
6
1
2
3
0
.
-
Dat a Entr y
AUTOCAL MODULE 36581KKF
40 MHz - 20 GHz
LEFT
RIGHT
Power
Operate
AUTOCAL
AUTOCAL
POWER
SUPPLY
37XXXC
SERIAL
PORT
SERIAL
CABLE
Figure 10-2.
AutoCal Equipment Setup
Содержание 37 C Series
Страница 3: ......
Страница 4: ......
Страница 18: ...Figure 1 1 Model 37XXXC Vector Network Analyzer System...
Страница 32: ......
Страница 41: ......
Страница 61: ...Intentionally Blank...
Страница 96: ...Table of Contents 5 1 INTRODUCTION 5 3 5 2 ERROR MESSAGES 5 3 Chapter 5 Error and Status Messages...
Страница 97: ......
Страница 103: ......
Страница 117: ...HARD COPY AND DISK OUTPUT DATA DISPLAYS 6 16 37XXXC OM Figure 6 12 Example of a Tabular Printout 37347C...
Страница 223: ......
Страница 242: ...37XXXC OM 9 21 9 22 TIME DOMAIN MENU FLOWCHART Figure 9 25 Time Domain Menu Flow...
Страница 276: ......
Страница 323: ......
Страница 567: ......
Страница 582: ...37100C 37200C 37300C Vector Network Analyzers Technical Data Sheet Vector Network Analysis up to 65 GHz...