
Using the 2300-584-R Software with the ME7828A
6-12 Restart for Alternate Frequency Band
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
PN: 10410-00286 Rev. L
6-33
6-12 Restart for Alternate Frequency Band
After the
Matched Thru
and
Mismatched Thru
tests have been successfully completed and the data saved, it is
necessary to restart the program and test the other frequency band. For example, if you tested for
40 MHz to
65 GHz Frequency Range
on your first calibration pass, on the second pass, select the
65 GHz to 110 GHz
Frequency Range
.
Procedure
1.
Ensure that the verification results of the previous test have been saved to the
C:\mmdscvc\Data
.
2.
To perform the verification tests for the second frequency range, on the
System Verification Software Main
Menu - Main Menu Tab,
select the
Restart
button. The main menu is described in
Interface – Main Menu Tab” on page 6-22
3.
The
Choose The Restart Method
dialog box appears as shown below in
.
4.
If you selected
Full Restart
, the calibration process starts again and the procedure steps are the same as
the procedures described previously starting in
Section 6-4, "Running the PVS Program"
,
5.
On the
Setup Menu
, repeat the data entry for the following items. Click
OK
when ready to proceed to the
next step.
6.
Enter the
GPIB Board 0.
7.
Enter the
GPIB Address 6
.
8.
Enter the
Test Set Serial Number.
9.
Enter the
User Name
.
10.
Enter the
Cal Kit Type
selection.
11.
Enter the
Cal Kit Serial Number
selection.
Figure 6-49.
Choose The Restart Method Dialog
Содержание 3656B W1
Страница 2: ......
Страница 12: ...Contents 10 PN 10410 00286 Rev L 3656B W1 Cal Ver Kit 2300 584 R PVS UG...
Страница 277: ......