Removal and Replacement
3-2
Battery Information
S312D MM
PN: 10580-00195 Revision B
3-3
Battery Testing Procedure
1.
With the Site Master
Off
and the battery installed, connect the Universal AC Adapter to the
12.5–15VDC (1350 mA) connector. The External Power LED and the Battery Charging LED will light.
2.
Disconnect the AC-DC Adapter when the Battery Charging LED turns off (indicating that the battery is
fully charged).
3.
Press and hold the
ESCAPE/CLEAR
key, then press the
ON/OFF
key to turn on the Site Master. This sets
the instrument to the factory preset state. Press
ENTER
when prompted to continue.
4.
Press the
SYS
key, followed by the
STATUS
soft key. Verify that the indicated battery charge is
≥
80%. If
the value is 80% or above, then press the
ESCAPE/CLEAR
key and continue with this procedure. If the
value is lower than 80%, then a discharge/charge cycle may be needed in order to improve the battery
capacity. Completely discharge the battery, as described in Step
and Step
below, and then recharge
the battery as described in Step
and Step
. If the battery capacity does not increase after a discharge/
charge cycle, then replace the battery.
5.
Press the
START CAL
key (to keep the Site Master from going into HOLD mode) and make note of the
test start time.
6.
When the Site Master display fades and the Site Master switches itself off, make note of the test stop
time.
7.
The total test time (Step
to Step
≥
1.5 hours. If the battery charge started at 80% or more,
and if the total battery test time is < 70 minutes, then replace the battery.
Note
If the Battery Charging LED does not light, the battery may be too low to immediately begin full
charging. Leaving the instrument connected to AC power for several hours may bring the battery up
to a level at which full charging can begin. Turn the instrument off and back on to see if the Battery
Charging LED lights indicate that a full charge cycle has begun.
Charging is inhibited below 0
°
C and above 45
°
C. If the instrument is too hot, the battery will not start
charging until the instrument temperature has cooled to 43
°
C.
Содержание Site Master S312D
Страница 2: ......
Страница 10: ...1 4 Replaceable Parts General Information 1 4 PN 10580 00195 Revision B S312D MM ...
Страница 48: ...3 7 RTC Battery Information Removal and Replacement 3 8 PN 10580 00195 Revision B S312D MM ...
Страница 58: ...Test Records A 10 PN 10580 00195 Revision B S312D MM ...
Страница 60: ...Test Fixture Schematics B 2 PN 10580 00195 Revision B S312D MM ...
Страница 61: ......