Cleaning the water filter 

To clean the water filter (located on the water supply line), follow the procedure detailed below. 

1.

 

 Unscrew the filter from the instrument. 

 

2.

 

 Unscrew the filter clamp from the water supply line. 

3.

 

 Use  compressed  air  to  blow  backwards  through  the 
filter  (blow  opposite  the  arrow  on  the  filter).  If  the 
filter is clogged, replace it with ANKOM part # X97. 

4.

 

 Reconnect the filter and fittings. 

5.

 

 Turn on the water and check for leaks. 

   

Replacing the Fuses 

To replace the fuses in the ANKOM

XT15

 Extraction System, follow the procedure detailed below. 

1.

 

 Turn off the instrument power and unplug the Power Cord from the outlet. 

2.

 

 Remove the two screws on the Top Cover. 

3.

 

 Swing the Top Cover down exposing the electrical compartment. 

4.

 

 Locate  the  Fuse  Holders  on  the  terminal  strip.  There 
are two fuses within the instrument; one for each pole 
(100 – 120V 10A or 220 – 240V 5A). 

 

  

5.

 

 Press the tabs to lift the Fuse Holders up. 

 

6.

 

 Replace the fuses and close the Fuse Holders completely. 

7.

 

 Close the Top Cover and secure it with the two screws that you previously removed. 

Replacing the Charcoal Filter 

The Charcoal Filter should be replaced every 3 months, or after 100 extraction operations using the instrument. To 
replace the filter, remove the old one from the vent line and install the new one. 

Fuse 
Holders 

Tab to open 
Fuse Holder 

Fuse  

Filter 
 
 
 

Содержание XT15

Страница 1: ...2052 O Neil Rd Macedon NY 14502 Telephone 315 986 8090 Fax 315 986 8091 www ankom com Operator s Manual Rev F 5 9 17 ...

Страница 2: ...This page intentionally left blank ...

Страница 3: ...ription 7 Safety Precautions 8 Instrument Installation 9 Fat Extraction Calculations 10 Fat Extraction Support Items 10 Sample Preparation 11 Fat Extraction step by step procedure using the ANKOMXT15 17 Periodic Maintenance 21 Troubleshooting Replacement Parts 23 Diagnostics 24 Appendix A Electrical Diagram pg 1 of 3 29 ...

Страница 4: ...This page intentionally left blank ...

Страница 5: ...filter bag handling as well as the instrument controls helping you to achieve the best possible results Please review the entire contents of this manual before you begin operating this product Warranty ANKOM Technology warrants the ANKOMXT15 Extraction System against any defects due to faulty workmanship or material for one year after the original date of purchase This warranty does not include da...

Страница 6: ...therefore always available to help you get the most from your ANKOM products We are also very interested in any comments or suggestions you may have to help us improve For any questions or suggestions regarding your instrument please contact us at Telephone 315 986 8090 Fax 315 986 8091 Email service ankom com www ankom com ...

Страница 7: ...icient porosity to permit rapid solvent passage and is composed of polymeric material that is resistant to the higher temperatures and solvents used in the instrument The quantitative isolation achieved by the filter bag permits the ANKOMXT15 Extraction System to automatically process samples in batches of up to 15 With the solvent reservoir filled simply place filter bags with sample in the Extra...

Страница 8: ...olatiles will not escape the vent system We recommend replacement of the filter every 3 months or after 100 extraction runs Trace amounts of vapor may also be present when samples are removed It is recommended that the ANKOMXT15 Extraction System be placed in a well ventilated environment away from any heat or ignition sources Organic solvents may be flammable Follow safe laboratory practices acco...

Страница 9: ... less than 1ml of volatiles It is the user s responsibility to determine whether venting of this instrument requires placement in a hood Please follow both local and federal regulations when installing this instrument 2 Connect the tube from the top fitting to the water supply and the one from the bottom fitting to the drain Ensure that there are no kinks or bends in any of the tubing connected to...

Страница 10: ... bag and Diatomaceous Earth DE after hydrolysis W3 Weight of dry extracted sample filter bag and DE C1 Blank filter bag dry weight after hydrolysis C2 Blank filter bag weight after extraction The Hydrolysis procedure can be found in the ANKOM HCl Operator s Manual Fat Extraction Support Items The following support items are needed to perform the fat extraction procedure Item Recommended Product El...

Страница 11: ...empty filter bag and the holder together 4 Add 1 5 2g of sample to the filter bag Keep all particles away from the sealing area of the filter bag 5 Record the weight of the sample W1 6 Set the Heat Sealer dial to 6 The setting may vary from sealer to sealer 7 Seal the filter bag within 4mm of its open end Keep the sealer arm down for 2 3 seconds after the red sealer light turns off to cool the sea...

Страница 12: ...rder to calculate the moisture value for the samples Moisture 100 x Filter Bag Weight Sample Weight Weight after drying Tin Weight Sample Weight Measuring moisture can also help verify the operation of your drying oven 2 Dry all weigh tins in an oven at 102 C 2 for 15 minutes 3 Remove weigh tins from the oven and place them in a Weigh Tin Desiccator or in a Desiccant Pouch 4 Allow weigh tins to co...

Страница 13: ...ags in their corresponding weigh tins for pre drying 14 Dry all samples at 102 C 2 for 3 hours to remove moisture prior to the extraction 15 Remove the weigh tins and filter bags from the oven and place them together in a Weigh Tin Desiccator A Weigh Tin Desiccator is recommended because it will hold between 10 and 14 weigh tins with filter bags depending on the size of the weigh tins A Desiccant ...

Страница 14: ...f the empty filter bag and the holder together 4 Add Diatomaceous Earth DE 4 1 Add 1 0 1 25 g of DE to the filter bag 4 2 Using a pencil eraser or similar sized utensil make an indentation or a space in the DE into which the liquid sample can be added using a pipette 4 3 Record the weight and tare 5 Pipette 1 5 2 g of liquid sample into the DE indentation in the filter bag Do NOT use more liquid s...

Страница 15: ... the extraction 11 Remove the samples from the oven and place them in a Desiccant Pouch 12 Allow the samples to cool to room temperature This should take about 10 15 minutes 13 Re weigh each bag W2 immediately after removing from the Desiccant Pouch The samples are now ready for the fat extraction procedure ...

Страница 16: ...Operator s Manual pg 16 Rev F 5 9 17 This page intentionally left blank ...

Страница 17: ...lay 2 2 Fill the solvent reservoir as needed 2 21 Unscrew the Fill Cap 2 22 Pour solvent into the Fill Port up to the dark blue level mark on the Solvent Sight Level About 500 ml of solvent is needed to reach the dark blue level mark Do not add solvent up to the top of the Solvent Sight Level tube or it will start flowing into the Vent Flask If using Diethyl Ether or a solvent mixture that include...

Страница 18: ...ction Vessel It is not necessary to remove the white PTFE Insert from the Extraction Vessel in order to put the Bag Holder in place 3 4 Put the Bag Holder with up to 15 bags into the PTFE Insert 3 5 If you removed the PTFE Insert put it back into the Extraction Vessel aligning the PTFE Insert slot with the Extraction Vessel pin The PTFE Insert O ring and the Extraction Vessel O ring must be in pla...

Страница 19: ...commended time is 60 minutes 4 2 Press ENTER on the Keypad to set the Time The Display should now show the screen to the right The temperature shown will be the last temperature entered 5 Set the Extraction Temperature 5 1 Using the arrow keys on the Keypad select the desired Extraction Temperature The recommended maximum temperature is 90 C 5 2 Press ENTER on the Keypad to set the Temperature The...

Страница 20: ...TFE Insert from the Extraction Vessel 8 4 Remove the Bag Holder from the PTFE Insert 8 5 Remove the filter bags from the Bag Holder 8 6 Dry the samples for 15 30 minutes in the oven at 102 2 C or in the ANKOMRD Dryer set at the factory setting of 110 C 8 7 Remove the samples from the oven or dryer and place them in a Desiccant Pouch 8 8 Allow the samples to cool to room temperature This should tak...

Страница 21: ...essel When cleaning the vessel do not damage the level sensor tip The tip may be cleaned with window cleaner or alcohol Cleaning the exterior of the instrument Wipe the outside of the instrument with window cleaner or a damp sponge as necessary Do not use Acetone to clean the instrument Cleaning the solvent filter If filter bags are not sealed properly and sample is lost from the bag it will be re...

Страница 22: ...iled below 1 Turn off the instrument power and unplug the Power Cord from the outlet 2 Remove the two screws on the Top Cover 3 Swing the Top Cover down exposing the electrical compartment 4 Locate the Fuse Holders on the terminal strip There are two fuses within the instrument one for each pole 100 120V 10A or 220 240V 5A 5 Press the tabs to lift the Fuse Holders up 6 Replace the fuses and close ...

Страница 23: ...s The ANKOM technology web site has the most current troubleshooting and replacement parts information Therefore if you have any questions about the operation of your ANKOMXT15 Extraction System or if you need replacement parts please visit our web site at www ankom com ...

Страница 24: ...Operator s Manual pg 24 Rev F 5 9 17 This page intentionally left blank ...

Страница 25: ...e Keypad four times The screen to the right will be displayed From the Diagnostics Screen you can access all of the available diagnostics Toggle Outputs To toggle outputs follow the procedure below 1 With the Diagnostics screen displayed press the ENTER button on the Keypad The screen to the right will be displayed If you press the STOP button on the Keypad while this screen is displayed you will ...

Страница 26: ...ss ENTER to get back to the main Diagnostics screen 6 Press the STOP button on the Keypad to go back to the main Outputs screen shown to the right 7 Press the STOP button on the Keypad to go back to the main Diagnostics screen shown to the right 8 Press the Up Arrow button on the Keypad to exit Diagnostics Press the Up Arrow button on the Keypad up to three more times to return to your solvent scr...

Страница 27: ...s not covered by liquid the Level Sensor should read 10 00v If the reading is less than 10 00v and the Level Sensor is not covered by liquid clean the Level Sensor Tip inside the Vessel with a cotton swab and alcohol If the Level Sensor reading is between 5 00 10 00v after cleaning the tip the instrument will run but you should consider replacing the tip Contact ANKOM Technology for more informati...

Страница 28: ...ght will be displayed 4 Press the Down Arrow button v on the Keypad The screen to the right will be displayed 5 Press the ENTER button on the Keypad The screen to the right will be displayed showing the state of the ETS relay If the ETS relay is OFF contact ANKOM Technology 6 Press the STOP button on the Keypad to go back to the Inputs screen shown to the right 7 Press the STOP button on the Keypa...

Страница 29: ...Operator s Manual Rev F 5 9 17 pg 29 Appendix A Electrical Diagram pg 1 of 3 ...

Страница 30: ...Operator s Manual pg 30 Rev F 5 9 17 Appendix A Electrical Diagram pg 2 of 3 ...

Страница 31: ...Operator s Manual Rev F 5 9 17 pg 31 Appendix A Electrical Diagram pg 3 of 3 ...

Страница 32: ...hod AOCS Am 5 04 Fully automatic Solvent recovery at 97 or greater Batch process up to 15 samples at one time RF Gas Production System High sensitivity pressure measurement Applications include Ruminant Nutrition Human Digestion Yeast Activity Beer Wine Fermentation Biomass to Energy analysis e g Ethanol Biodegradability Soil respiration BOD etc Wireless Computer control and data storage Chemicals...

Отзывы: