Periodic Maintenance
Initial Maintenance (after 10 hours of operation)
After the first 10 hours of operation, follow the procedure below:
1.
Remove the side window of the instrument.
Side view of instrument with door removed
2.
Inspect the Maintenance Alert Collector.
3.
Clean any residue from the collector.
IMPORTANT:
Do NOT insert the Bag Suspender into the instrument for this procedure.
4.
Press the arrow keys on the Keypad until you see “Diag Mode” on
the Display. Press “ENTER”.
5.
Use the arrow keys to scroll until you see
Q5 Agitator
on the screen. Press “ENTER”.
6.
When the motor is activated, turn the Packing Nut to
the
RIGHT
until you hear a change in the sound of the
motor. (The motor will start to labor as the packing nut
gets harder to turn.)
7.
Loosen the Packing Nut slightly until the motor stops
laboring.
8.
Turn off the instrument and re-install the side window.
Repeat the Initial Maintenance procedure (steps 1 – 10 above) every month of operation (120 hours) for the first 3
months. After that, execute the procedure once every 2 months.
Diag Mode
v
<Enter>
Packing Nut
Maintenance Alert Collector
Содержание Delta
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 18: ...Operator s Manual pg 18 Rev 4 24 20 This page intentionally left blank ...
Страница 24: ...Operator s Manual pg 24 Rev 4 24 20 This page intentionally left blank ...
Страница 30: ...Operator s Manual pg 30 Rev 4 24 20 This page intentionally left blank ...